Publications Database

HexoskinAbstract & PosterCardiology & ArrhythmiaRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceClinical Medicine & SurgeryValidation & Methodology

HexoskinAcademic ThesisData Science & Artificial IntelligenceValidation & Methodology

Key FindingIn the gait experiment, the eyeglass-camera system significantly discriminated gait speed and outperformed the vest-based accelerometer, while motion capture provided superior gait-speed discrimination. The glasses-derived frontal-plane data distinguished slow, medium, and fast speeds in all reported pairwise comparisons (all p < .05); the accelerometer's vertical-plane data did not (p = .26). Separately, a video-based Fuzzy Inference System reported 74% overall activity-classification accuracy, but the dissertation cautions against direct comparison with systems that produce one definitive action label.

Summary

In the gait experiment, 19 healthy college students aged 18–21 years (10 male and 9 female) completed six over-ground walks covering a distance of four meters at self-determined slow, medium, and fast speeds in a motion-capture laboratory. Hexoskin supplied tri-axial acceleration at 64 Hz; its other sensor data were not analyzed. The experiment compared the eyeglass-camera system and the body-worn accelerometer with 20-camera optical motion capture for classifying gait speed. Separately, the dissertation describes a video-based Fuzzy Inference System that uses video features and identified objects to classify activities.

HexoskinArticleExercise Physiology & Sports ScienceValidation & Methodology

Key FindingDuring a 530-second lead climb monitored with Hexoskin, heart rate consistently increased during active climbing and decreased during rest, while breathing rate varied less consistently. Hip-acceleration peaks up to 0.36 g corresponded to active climbing, and sustained values of 0.01–0.08 g corresponded to rest. In a difficult 46–151 s section, heart rate, breathing rate, and minute ventilation all rose substantially; these descriptive results support the feasibility of linking sensor data with climbing video, and they do not show that a training intervention improved performance.

Summary

This study developed a method for examining time-resolved biometrics in relation to climbing strategy, technique, and position on an indoor wall. Continuous Hexoskin measurements and smartphone video were time-matched so heart rate, breathing rate, minute ventilation, and hip acceleration could be viewed at specific positions along the route. An interactive interface supported review of biometric changes at key events and calculation of summary statistics. The illustrated demonstration was a 530 s (8:50, min:sec) lead climb on a 16.8 m (55 ft) indoor wall.

HexoskinArticleClinical Medicine & SurgeryData Science & Artificial Intelligence

Key FindingFor activity recognition from Hexoskin acceleration, the multivariate CNN and the three-pipeline ensemble each reported weighted-average precision, recall and F1 scores of about 99% across their evaluated activity classes. The ensemble covered ten activities, and under the reported setup, parallel processing made segmentation almost five times faster while GPU execution made the CNN about 80 times faster than CPU execution.

Summary

Abid and colleagues investigated inter-subject human-activity recognition from acceleration data collected with a Hexoskin shirt. They compared three classification pipelines based on handcrafted features with ReliefF and SVM, LDA with KNN, and a multivariate CNN, then combined them in an ensemble for ten static, dynamic and hazardous activities. The ensemble reported weighted-average precision, recall and F1 scores of about 99%, while parallel processing shortened segmentation and GPU execution accelerated the CNN under the reported setup.

HexoskinArticleCardiology & ArrhythmiaValidation & Methodology

Key FindingProcessed waist ECG generally tracked the Hexoskin chest-reference heart rate during walking and jogging, but tracking deteriorated during two approximately five-minute poor-quality periods. Artifact cancellation was less complete during walking than jogging. A secure waistband fit remained necessary because the method could not recover QRS complexes fully obscured by motion artifact; results come from one participant, and RMSSD agreement applies only to acceptable-quality segments.

Summary

This single-participant preprint evaluated a peak-space motion-artifact cancellation method for waist textile ECG during outdoor walking and jogging. A 54-year-old man completed four sessions on different days, each with approximately 20–25 minutes of walking and jogging. Hexoskin chest ECG served as the reference for waist ECG recorded with a Zephyr Bioharness.

HexoskinArticleNeuroscience, Sleep & CognitionOccupational Health & Human FactorsData Science & Artificial Intelligence

Key FindingInterViewR combined immersive virtual-reality interview practice with Hexoskin-based affective feedback, and the authors reported that it elicited stress responses and provided coherent feedback on interview performance. In a separate 11-person usability evaluation, none of the participants had Autism; they considered the system useful for interview training but did not expect to use it frequently. Suggested improvements included a more human-like interviewer, more varied settings, clearer feedback and objective performance scoring.

Summary

Ahmed and colleagues developed InterViewR to address limitations in digital interview training for individuals with Autism by combining an immersive virtual-reality interview with a Hexoskin smart shirt. Hexoskin captured RR intervals and breathing rate so the platform could present real-time affective feedback and help users recognize stress during different parts of an interview. The user evaluation examined feasibility, usability, and perceived usefulness, and the authors reported that the mixed-reality platform elicited genuine stress responses while providing coherent feedback on interview skills.

HexoskinArticleExercise Physiology & Sports Science

Key FindingHeart rate measured with Hexoskin was 15–20 bpm lower at each power stage with Atenolol, while DFA a1 declined similarly with increasing cycling power in the control and Atenolol trials. At roughly 125 bpm, DFA a1 was 0.336 during Atenolol cycling and 0.723 during the control ramp at different workloads. These observations distinguish matching heart rate from matching workload in this individual; they do not establish universal training thresholds.

Summary

This single-participant case report examined DFA a1, a nonlinear heart-rate-variability index, during cycling in a 62-year-old male recreational cyclist. It compared progressive five-minute intervals with and without 25 mg Atenolol and longer 10-minute intervals near the first ventilatory threshold. Hexoskin supplied heart-rate and R–R interval recordings, which were analyzed alongside ventilation and cycling workload.

HexoskinArticleCardiology & ArrhythmiaExercise Physiology & Sports Science

Key FindingIn ten professional golfers, olfactory inhalation did not significantly change putting success (p = 0.115) or heart rate recorded by Hexoskin during the address, putting, or total phases (p = 0.838, 0.878, and 0.666). Postural stability measured by centre-of-pressure area was significantly inhibited in the address, downswing, and total putting sections (p = 0.000, 0.035, and 0.047).

Summary

Ten professional golfers completed 3-metre putting tasks under control, lavender, citrus, and ammonia conditions. A Hexoskin biometric shirt recorded heart-rate changes, while a force plate measured centre-of-pressure area. Olfactory inhalation produced no significant difference in putting success or heart rate, but postural-stability differences were significant in the address, downswing, and total putting sections.

HexoskinArticleCardiology & ArrhythmiaClinical Medicine & SurgeryNeuroscience, Sleep & Cognition

Key FindingThis yoga therapy protocol used Hexoskin to monitor participants after cancer treatment; it reports no efficacy results. It describes a single-subject design with 25 recruited adults and six dropouts, using heart-rate variability as the primary outcome and fatigue, anxiety, depression, stress, perceived cognitive function and quality of life as secondary outcomes.

Summary

This protocol describes an exploratory single-subject study of yoga therapy after cancer treatment. Twenty-five adults were recruited and six dropouts were reported. The program comprised one individual yoga-therapy session and six weekly group sessions, with seven assessments from baseline through follow-up. Hexoskin shirts recorded ECG and respiratory data at assessment visits, and hierarchical linear modeling was planned for heart-rate variability and patient-reported outcomes. The publication reports a protocol rather than efficacy results.

HexoskinAcademic ThesisRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceClinical Medicine & SurgeryValidation & Methodology

Key FindingIn five analyzable participants with COPD, Hexoskin tracked changes in breathing pattern during stair stepping, but the comparison with Vmax did not support using the shirt to detect bronchodilator-induced changes in operating lung volumes.

Summary

Six adults with chronic obstructive pulmonary disease completed stair-stepping tests before and after bronchodilator treatment while wearing Hexoskin. Researchers compared the shirt with a Vmax metabolic cart; poor shirt signal quality excluded one participant from the main analysis. Tidal volume, minute ventilation and respiratory rate correlated between devices, but inspiratory capacity and inspiratory reserve volume did not. Changes after bronchodilator treatment did not correlate significantly between devices.

HexoskinArticleClinical Medicine & SurgeryNeuroscience, Sleep & Cognition

Key FindingThe authors reported that cancer-related fatigue and state anxiety increased immediately after the individual session but improved more during the subsequent group-intervention phase. Trait anxiety, perceived cognitive impairment, its impact on quality of life, and functional wellbeing also improved over time, while the remaining patient-reported outcomes did not change. These exploratory findings require confirmation in controlled trials and do not provide garment-derived physiological outcomes.

Summary

This exploratory single-subject study analyzed 20 of 25 enrolled adults after cancer treatment. Participants received one individual yoga-therapy session and six weekly small-group sessions, with patient-reported fatigue, anxiety, cognitive function, depression, stress, and quality of life measured across phases. Hierarchical linear modeling compared changes over time. The article reports preliminary patient-reported outcomes rather than garment-derived physiological findings.

HexoskinArticleOccupational Health & Human Factors

Key FindingIn a study pairing questionnaires with Hexoskin heart-rate variability monitoring, educators reported poorer functioning than normative data for emotional role limitations, energy/fatigue, and social functioning (BF10 > 100); this self-report comparison does not establish a physiological burnout marker. The mental-health association with RMSSD was small (τb = 0.17, p < .05) with anecdotal Bayesian support (BF10 = 2.25). Work–family conflict associations with RMSSD and HF power became nonsignificant after controlling for age, and the paper reported evidence against a relationship between burnout scores and physiological measures.

Summary

This exploratory study combined psychological surveys with Hexoskin HRV monitoring in Australian early childhood educators. One hundred educators were recruited, and 68 had usable physiological data after cleaning. A common four-hour window from 10:00 a.m. to 2:00 p.m. was analyzed across work shifts; it was not a uniform total wear duration.

HexoskinConference AbstractCardiology & ArrhythmiaClinical Medicine & SurgeryValidation & Methodology

HexoskinArticleCardiology & ArrhythmiaExercise Physiology & Sports Science

Key FindingPM2.5 and PM10 doses estimated using environmental sensors and Hexoskin minute ventilation were highly correlated (r = 0.98), so the authors retained PM2.5 for subsequent analysis. Separate models reported associations of higher cumulative PM2.5 and noise dose with greater heart-rate variation, but the noise association disappeared when PM2.5 entered the combined model. The authors reported no HRV effect and cautioned against generalizing from four participants; the source's Bayes-factor interpretation is internally inconsistent, so no evidence-strength direction is assigned here.

Summary

This exploratory study followed four healthy young cyclists aged 25–37 years in Montreal over five days to examine traffic-related particulate-matter and noise exposure alongside pre- and post-cycling cardiac measures. Hexoskin supplied continuous ECG during 40-minute sitting periods and minute ventilation used with Aeroqual pollutant concentrations to estimate inhaled dose. Bayesian models analyzed 255 five-minute RR-interval segments.

AstroskinArticleCardiology & ArrhythmiaClinical Medicine & SurgeryValidation & Methodology

Key FindingAstroskin heart rate and systolic blood pressure showed low error, strong agreement with corpuls3, and comparable reliability across sessions. SpO2 error was low, but concordance and reliability were weaker, potentially reflecting limited variability and sensor placement.

Summary

MacQuarrie and colleagues compared Astroskin with the hospital-grade corpuls3 monitor in 24 participants during a seated 22-minute session; 19 returned for a second seated session. Heart rate and systolic blood pressure showed low error, strong concordance, and comparable between-session reliability. Oxygen-saturation error was low, but concordance and reliability were weaker, identifying SpO2 sensor placement and performance across a wider physiological range as areas for further study.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingIn five surgeons monitored with Hexoskin across 179 joint replacements, energy expenditure was descriptively higher during conversion/complex primary and revision arthroplasty than during simple primary procedures. After accounting for surgeon differences and patient BMI, complex primary knee arthroplasty remained higher than simple primary knee arthroplasty (β = 166.2, SE 60.7; P = 0.0073), whereas the corresponding hip comparison was not significant (P = 0.17). Heart rate and minute ventilation showed no significant overall group differences.

Summary

Five orthopaedic surgeons wore Hexoskin garments during 179 joint-replacement operations to examine physiological workload across simple primary, conversion/complex primary, and revision knee and hip arthroplasty. The garment-derived outcomes were heart rate, minute ventilation, and energy expenditure. The patient case counts and demographics describe the operations, not the five monitored surgeons.

HexoskinAcademic ThesisExercise Physiology & Sports ScienceOccupational Health & Human FactorsValidation & Methodology

Key FindingThe safety model estimated that adding Hexoskin and infrared cameras would reduce the simulator's failure-mode detection score to 4 and its risk-priority number from 144 to 72. This was a modeled safety-analysis result rather than a prospective validation of Hexoskin with firefighter trainees.

Summary

Sherman's system-safety study examined a crawling training device used by East Alabama fire departments, where reducing training-related injury risk was the central objective. The analysis proposed Hexoskin for remote physiological monitoring alongside infrared cameras and modeled it as a failsafe in the simulator fault tree. The model estimated that these additions would reduce the failure-mode detection score to 4 and the risk-priority number from 144 to 72; this was a modeled result, not a prospective validation with firefighter trainees.

HexoskinArticleOccupational Health & Human FactorsValidation & Methodology

Key FindingOf five interface gestures based on Hexoskin chest and abdominal movement, breath holding and thoracic sighing received the highest ease ratings. In a separate 'worked well' question, five of seven participants selected hold breath and five selected deep thoracic breath. Several gestures required more than one detection attempt, and participants reported difficulty distinguishing thoracic from abdominal breathing; the authors identified blurred gesture distinctions and repeated attempts as reasons for further refinement and personalized calibration; these were future work, not demonstrated improvements.

Summary

Using Hexoskin thoracic and abdominal respiratory movement, this preliminary study developed five breathing gestures as inputs for a game-like Android application. Seven users tested the gestures and rated the experience an average 9 out of 10. The study identified the need for more personalized detection and calibration to accommodate individual breathing patterns.

HexoskinArticleCardiology & ArrhythmiaRespiratory & Infectious DiseaseValidation & Methodology

Key FindingNon-contact radar estimates correlated 94% with Hexoskin breathing rate and 80% with Hexoskin heart rate. Hexoskin functioned as the wearable reference sensor, so the reported correlations evaluate the radar estimates rather than establish a new Hexoskin accuracy estimate.

Summary

Alizadeh and colleagues used a 77 GHz frequency-modulated continuous-wave radar in a bedroom setting to estimate breathing and heart rates from a person lying on a bed. Hexoskin served as the wearable reference sensor for both measures. Radar estimates correlated 94% with Hexoskin breathing rate and 80% with Hexoskin heart rate under the reported conditions, so these results evaluate the radar rather than establish a new Hexoskin accuracy estimate.

HexoskinAcademic ThesisOccupational Health & Human Factors

Key FindingAmong RCMP law-enforcement students, greater intolerance of uncertainty was associated with lower combined driving performance in the second session. Intolerance of uncertainty was also positively associated with task appraisal, while subjective stress was not significantly associated with the reported performance scores or maximum heart rate. These are correlational findings and do not show that either construct caused changes in stress or performance; the paper's abstract gives a broader null conclusion that conflicts with the detailed driving result, so the mixed findings should be interpreted cautiously.

Summary

This study examined how intolerance of uncertainty and task appraisal relate to stress and performance in RCMP law-enforcement students. Participants completed questionnaires and simulated emergency-driving and use-of-force tasks. Hexoskin recorded heart rate as a measure of physiological arousal.

HexoskinArticleRespiratory & Infectious DiseaseNeuroscience, Sleep & CognitionClinical Medicine & SurgeryValidation & Methodology

Key FindingUsing Hexoskin as the breathing-rate reference, the contactless radar achieved 97% overall breathing-rate accuracy and a maximum RMSE of 1.7 bpm across the reported measurements. Sleep position was classified correctly in more than 83% of cases, supporting simultaneous breath and position monitoring of two nearby sleepers under the tested conditions.

Summary

Alizadeh and colleagues developed a contactless MIMO-radar system to monitor breathing and sleep position for two adjacent bedbound subjects at a time. The study reports three experimental rounds with two subjects per round, but it does not state that these were six distinct people. Hexoskin served as the breathing-rate reference. The radar reached 97% overall breathing-rate accuracy, a maximum RMSE of 1.7 breaths/min and more than 83% correct sleep-position detection under the reported conditions.

HexoskinArticleData Science & Artificial IntelligenceOccupational Health & Human Factors

Key FindingIn five undergraduates monitored with Hexoskin vests and Actigraph watches, a temporal random-forest model using a 30-day sliding training window reached about 92% median accuracy and F1. That result came from temporal cross-validation; person-specific models performed poorly when trained on one semester and evaluated on a later semester. The study did not establish performance in larger or more diverse groups; stress labels were self-reported, sensor coverage was limited, and factors such as sleep, caffeine and outside stress were not controlled.

Summary

Researchers used Hexoskin vests and Actigraph watches to collect physiological and movement data for classifying self-reported classroom stress in five undergraduates. They compared several machine-learning approaches, including models that preserved the timing of observations. A temporal random-forest analysis with a 30-day sliding training window reported about 92% median accuracy and F1, but performance was weaker when person-specific models were tested on a later semester. The small, demographically narrow sample limits generalization.

HexoskinArticleNeuroscience, Sleep & CognitionData Science & Artificial Intelligence

Key FindingWith multimodal inputs including Hexoskin ECG, respiration and acceleration, XGBoost reached 0.95 accuracy and 0.94 F1 score on the combined calibration subset and 0.92 accuracy and 0.90 F1 score on WEAR in-lab data. In transfer learning, Random Forest performed best for calibration and in-lab data, while SVM led for in-the-wild data, so the baseline leaders did not consistently retain their advantage. The RADWear in-the-wild data were 80% not anxious, and the researchers undersampled that majority class while preserving the minority observations.

Summary

This study evaluated whether machine-learning models developed with wearable data from controlled conditions could extend to real-world anxiety detection in young adults. Hexoskin contributed ECG, respiration, and 3-axis acceleration data within a multimodal system that also used an E4 wristband. XGBoost and Decision Tree performed strongly when models were trained and tested on the same subsets, whereas Random Forest led transfer learning in calibration and in-lab settings and SVM led for in-the-wild data. The findings show that environmental noise, multimodal signal contributions, and the balance of feature importance affect model robustness and transferability.

HexoskinAcademic ThesisExercise Physiology & Sports ScienceOccupational Health & Human FactorsValidation & Methodology

Key FindingDuring 232 monitored shifts, the 32-paramedic field cohort spent more time on calls per shift in metropolitan postings than in regional postings. In a separate 14-paramedic fitness substudy, greater estimated aerobic fitness was associated with lower heart rate at shift start and across shifts. These observational results do not establish causation, and the fitness subgroup was self-selected; a cardiac-arrest analysis was limited to four paramedics and seven arrests.

Summary

This thesis combined a health survey, field monitoring, a fitness substudy and laboratory validation to examine paramedic health and work demands. In the field study, Hexoskin recorded heart rate, respiratory rate, physical activity, energy expenditure and steps during 232 shifts completed by 32 paramedics. A separate laboratory study compared Hexoskin heart-rate and respiratory-rate measurements with standard measurements during cycling. Participant counts and results belong to distinct study components and should not be combined.

AstroskinAcademic ThesisExercise Physiology & Sports ScienceData Science & Artificial Intelligence

Key FindingActivity-recognition accuracy declined when wearable hardware or recording sessions changed: Model A fell from 99.07% on Astroskin data to 65.74% on BioHarness data, while Model A Type 1 fell from 99.57% to 50.95% across Astroskin sessions. The BioHarness-trained Model H achieved 29.63% on Astroskin data, but the thesis gives conflicting values for its same-domain BioHarness result; when both hardware and session differed from training, accuracy reached 19.28%. The out-of-domain discriminator flagged unfamiliar device or session data with 97.60% accuracy, but the experiment used only the author as a participant in a home setting after COVID-19 restrictions prevented the planned research, so broader performance remains unestablished.

Summary

This thesis evaluated how human-activity-recognition models performed when wearable hardware or recording sessions changed, using Astroskin and Zephyr BioHarness accelerometer data collected from the author as the sole participant in a home setting after COVID-19 restrictions prevented the planned research. An Astroskin-trained model dropped from 99.07% accuracy on Astroskin data to 65.74% on BioHarness data, while another model declined from 99.57% to 50.95% when tested in a different Astroskin recording session. A separate out-of-domain discriminator identified data from unfamiliar hardware or recording sessions with 97.60% accuracy. The thesis proposes this discriminator as a warning when incoming sensor data no longer match a model’s training conditions, but the one-person home experiment does not establish broader performance.

HexoskinArticleClinical Medicine & SurgeryNeuroscience, Sleep & Cognition

Key FindingIn a crossover trial that used Hexoskin for exploratory monitoring, five days of low-dose oral ketamine did not improve clinical outcomes over placebo in girls with Rett syndrome, although exploratory EEG findings showed first-dose target engagement. Most related adverse events were mild or moderate; one participant stopped treatment after Grade 3 vomiting; only 10 participants contributed to the EEG analysis, and no correction was applied for multiple comparisons. The trial tested only the two lowest planned doses and stopped after two cohorts because of recruitment difficulties, limiting conclusions about higher doses or longer treatment.

Summary

A randomized, double-blind crossover trial enrolled 23 girls aged 6–12 years with Rett syndrome. Participants received five days of oral ketamine or placebo, followed by a nine-day washout and the alternate treatment; the studied doses were 0.75 or 1.5 mg/kg per dose, given twice daily. Hexoskin contributed daily activity, sleep, heart and lung-function data as exploratory measures. The primary outcome was safety and tolerability.

HexoskinArticleCardiology & ArrhythmiaRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceClinical Medicine & SurgeryData Science & Artificial Intelligence

Key FindingAcross cycling intensities, a temporal convolutional network using 218 seconds of Hexoskin cardiorespiratory data produced a combined oxygen-uptake error of −22 ± 122 ml min−1. The temporal models had lower limits of agreement and substantially fewer parameters than the stacked LSTM comparison, while a heart-rate-only configuration showed high bias and variance. Predicted oxygen uptake classified light, moderate and vigorous activity with 94.1% accuracy across 16,090 seconds.

Summary

Amelard and colleagues used Hexoskin smart-shirt cardiorespiratory signals and reference oxygen-uptake measurements from a metabolic system in 22 healthy young adults completing cycle-ergometer protocols. A temporal convolutional network modeled dynamic oxygen uptake across exercise intensities. The predicted oxygen uptake classified light, moderate and vigorous activity second by second with 94.1% accuracy across 16,090 seconds. The authors describe the approach as a path toward quantitative aerobic-activity monitoring outside the laboratory when combined with tidal-volume and heart-rate-reserve calibration.

HexoskinArticleExercise Physiology & Sports Science

Key FindingDuring a roughly 17-minute race in which Hexoskin tracked heart rate and separate instruments measured temperature, nine male stock-car drivers lost 0.77 ± 0.3% of body weight on average and had a mean sweat rate of 0.63 ± 0.4 L/h. Core and skin temperatures rose, while the core-to-skin temperature gradient narrowed. The sweat-rate calculation included prerace activity, and the study did not measure the three- to four-hour fluid losses discussed as projections or test effects on driving performance.

Summary

Nine male stock-car drivers were monitored during a competitive NASCAR event in hot conditions. Hexoskin continuously recorded heart rate during the racing session, while separate instruments measured core and skin temperature and environmental conditions. The study described hydration and physiological strain during a short race; it did not test whether these changes affected driving performance.

HexoskinAcademic ThesisCardiology & ArrhythmiaRespiratory & Infectious DiseaseNeuroscience, Sleep & Cognition

Key FindingTwo young adults with ADHD completed the seven-week Hexoskin-supported biofeedback intervention. Both showed improved emotional regulation after the intervention and at the two-week follow-up, while emotional-awareness results varied by participant and instrument, with EDRE scores returning to baseline at follow-up. With two completers and no control group, the findings are preliminary and cannot establish that biofeedback caused the observed changes.

Summary

Côté-Bélisle explored Hexoskin-supported biofeedback for emotional regulation and emotional awareness in young men with ADHD. Participants used heart-rate and breathing-rate reports to identify physiological-arousal surges and connect them with everyday emotional experiences. Both completers later reported greater competence in identifying, understanding, and managing their emotions.

HexoskinArticleData Science & Artificial IntelligenceOccupational Health & Human FactorsValidation & Methodology

Key FindingIn the tested digital-twin setup, Bio IK used 38 body points and produced lower avatar-joint position error than Unity IK using nine points. Bio IK required more processor time but slightly less memory, and the authors integrated it into an augmented-reality application. The comparison concerns avatar positions measured against motion capture; it does not validate Hexoskin biometrics or establish that one algorithm performs better for every movement or system.

Summary

This prototype compared two inverse-kinematics algorithms for augmented-reality digital twins of manual-handling tasks. Motion-capture joint positions served as the reference for avatar-joint error. Participants also wore a Hexoskin Shirt and Smart Device to supply heart rate, breathing rate and calorie data for the broader fatigue-analysis setup, but the study did not report a biometric or fatigue-detection validation.

AstroskinPaperNeuroscience, Sleep & CognitionOccupational Health & Human FactorsClinical Medicine & Surgery

Innovative Biomonitoring Systems in the Aerospace Industry

Key FindingAstroskin is discussed as a potential source of physiological data. The paper reports a design proposal, with no new wearer measurements or clinical outcome.

Summary

This paper reviews physiological monitoring aboard the International Space Station and describes Astroskin in the context of a student proposal for a space medical center. The proposed center would link crew monitoring with physician support.

AstroskinPaperData Science & Artificial IntelligenceOccupational Health & Human Factors

Key FindingThe proposed service would move beyond downloading data after wear. The paper describes a system design; it reports no test of real-time Bio-Monitor transmission or astronaut health outcomes.

Summary

This engineering paper proposes a way to combine Bio-Monitor recordings with other astronaut health data and transmit them through space data relays. It describes how the service could support monitoring beyond low Earth orbit.

HexoskinAcademic ThesisNeuroscience, Sleep & CognitionClinical Medicine & Surgery

Key FindingAmong the 27 participants retained for momentary analyses, resting-state connectivity between the amygdala and ventromedial prefrontal cortex was associated with specific outcomes. Right-sided connectivity was associated with sadness instability and the behavioral-dysregulation composite, while binge eating was associated with connectivity on both sides; overall mean negative affect and overall affective instability showed no significant association. These are specific, non-causal fMRI and ecological-momentary-assessment associations: although participants wore Hexoskin during the 14-day awake data-collection period, the dissertation excluded its cardiac, respiratory and accelerometry data from analysis and reports no Hexoskin physiological results.

Summary

This dissertation examined associations between resting-state amygdala–ventromedial prefrontal cortex connectivity and emotional distress in daily life using functional MRI and 14 days of ecological momentary assessment. Thirty-two women completed the procedures; three left-handed participants were excluded from all analyses, leaving 29 right-handed women, and two participants with low random-prompt compliance were further excluded from momentary analyses, leaving 27. Right amygdala–ventromedial prefrontal cortex connectivity was associated with sadness instability and a behavioral-dysregulation composite, while overall mean negative affect and overall affective instability were not significantly associated with connectivity. Participants wore Hexoskin while awake to collect cardiac activity, respiration and accelerometry, but the dissertation explicitly excluded all physiological data from its analyses and reports no Hexoskin physiological results.

AstroskinAcademic ThesisExercise Physiology & Sports ScienceData Science & Artificial IntelligenceValidation & MethodologyCardiology & Arrhythmia

Key FindingAcross five participants in a daily-activity analysis, complete blood-pressure waveform estimates had a mean absolute error of 6.18 ± 0.73 mmHg when walking data were included and 5.85 ± 0.73 mmHg when they were excluded. Separate exercise analyses found that nonlinear autoregressive models with external inputs (NARX) tracked large mean arterial pressure changes, but did not capture changes linked to peripheral resistance from ECG and forehead PPG alone.

Summary

This thesis developed data-driven models that estimate the full blood-pressure waveform from electrocardiogram and photoplethysmography signals and attach an uncertainty measure to each estimate. Astroskin supplied electrocardiogram and photoplethysmography signals for both study components and tri-axial acceleration for daily-activity monitoring. The daily-activity study used Portapres as the blood-pressure reference, while the cycling study used NOVA from Finapres Medical Systems; reference-device failures excluded six of 18 cycling participants, leaving 12 for analysis. The studies used distinct participant groups, so their results remain tied to their individual protocols.

AstroskinPaperCardiology & ArrhythmiaOccupational Health & Human FactorsClinical Medicine & SurgeryData Science & Artificial Intelligence

Key FindingThe authors reported that generated signals were 3.6 times closer to the original asymptomatic wearable ECGs and 1.3 times closer to the symptomatic clinical ECGs than the baseline model’s outputs. This measured signal similarity; it did not test diagnosis in astronauts.

Summary

Researchers used Astroskin ECG data to test a method for generating synthetic signals that combine features of wearable recordings with patterns from clinical ECGs. The study compared the generated signals with those from a model trained on mean squared error alone.

HexoskinArticleData Science & Artificial IntelligenceCardiology & Arrhythmia

Key FindingThe support-vector-machine classifier reached 95.4% overall activity-classification accuracy under leave-one-subject-out testing. It used Hexoskin acceleration data from 40 healthy volunteers, not cardiac-rehabilitation patients. The application displayed recognized activities and included an archive of medical reports for doctors, but increased home-rehabilitation participation remained an aim rather than a demonstrated outcome.

Summary

This study developed a mobile human-activity-recognition application using three-axis acceleration from a Hexoskin garment. The classifier was evaluated on data from 40 healthy volunteers who repeated 10 activities and achieved 95.4% overall accuracy with a support vector machine under leave-one-subject-out testing. Home cardiac rehabilitation was the intended application, but no rehabilitation patients or clinical outcomes were evaluated.

HexoskinAcademic ThesisData Science & Artificial Intelligence

Key FindingThe Hexoskin evaluation set focused on sleep and walking annotations, and the classifiers were assessed with accuracy, recall, precision and F1 score. The LSTM without manual feature engineering produced the strongest balanced test result reported in the table: 98.4% accuracy, 97.5% recall, 93.4% precision and 95.4% F1 score. Several tree and boosting configurations showed train-test gaps consistent with overfitting, and the thesis does not establish effectiveness in healthcare settings.

Summary

Saini evaluated machine-learning and deep-learning models for human-activity recognition using participant-labeled Hexoskin data. Models trained with Capture24 were tested on Hexoskin data for activities including walking and sleeping. The conclusion reports 98.4% accuracy for an LSTM configuration that did not require manual feature engineering to distinguish walking from sleeping. The thesis presents healthcare use as a future possibility rather than a demonstrated clinical benefit.

HexoskinAcademic ThesisExercise Physiology & Sports ScienceOccupational Health & Human FactorsValidation & Methodology

Développement d’une veste de refroidissement individuelle adaptée au travail en mines profondes

Key FindingIn the cooling study, which used Polar H10 for heart rate, wearing the garment was associated with lower internal body temperature and heart rate, lower perceived exertion and better comfort, humidity sensation and well-being during the 42-minute test. In the separate validation, average Hexoskin–Polar heart-rate correlations were 0.9987 in moderate conditions and 0.9985 in hot, humid conditions. These correlations do not establish individual interchangeability, and no miner field deployment was tested.

Summary

This thesis contains two separate human studies. A 42-minute cooling-garment test involved 19 male completers and evaluated a liquid-carbon-dioxide system in a hot, humid chamber. A second study compared Hexoskin and Polar H7 heart-rate measurements in 12 male volunteers during a 28-minute cycling protocol under moderate and hot, humid conditions. Cooling outcomes came from separate Polar H10 and thermometer measurements.

HexoskinArticleExercise Physiology & Sports Science

Key FindingIn Delhi, researchers combined Hexoskin ventilation data with environmental measurements. Cycling on a primary road rather than a residential street was associated with 47 µg/m³ more NO2 and 3.3 dB(A) more noise. After one hour in areas with moderate predicted pollution, estimated noise dose reached 212% on residential streets and 459% on primary roads, while inhaled NO2 reached 403 µg and 482 µg, respectively.

Summary

Apparicio and colleagues monitored three healthy male cyclists over five dry days in Delhi while modeling road-level noise and nitrogen-dioxide exposure. Hexoskin measured heart rate, breathing rate and minute ventilation; minute ventilation was combined with Aeroqual nitrogen-dioxide concentration to estimate inhaled dose. For one hour of cycling under moderate predicted pollution, estimated noise dose was 212% on residential streets and 459% on primary roads, while estimated inhaled nitrogen dioxide was 403 µg and 482 µg, respectively.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human FactorsValidation & Methodology

Key FindingHeart rate measured with Hexoskin and cortisol measured in saliva rose more during real emergency-department resuscitations than during virtual simulations. Virtual-reality scenarios still increased heart rate, but the study did not establish equivalent physiological stress. This was a small, single-site provider study: nine shifts had entirely missing heart-rate data, and resuscitation type and patient outcomes were not collected; the real and virtual phases occurred in different settings and periods, so the comparison does not isolate simulation format alone.

Summary

A single-site pilot compared physiological stress in 16 pediatric emergency physicians during real emergency-department resuscitations and a one-hour virtual-reality session. Across 69 monitored shifts and 31 resuscitations, heart-rate and salivary-cortisol increases were larger in real events, while perceived workload was higher in virtual reality. Hexoskin measured heart rate; cortisol and workload were separate measures.

HexoskinAcademic ThesisRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceClinical Medicine & Surgery

Key FindingMost participants increased tidal volume before accelerating breathing frequency late in exercise, but one third of the asthma group increased breathing frequency earlier, at about 30% to 40% of maximum workload. Both markedly increasing and unusually stable thoraco-abdominal asynchrony were associated with signs of dysfunctional breathing. These deviant patterns occurred in 33% of participants with asthma and 7% of healthy controls.

Summary

Pelgröm used Hexoskin thoracic and abdominal respiratory-inductance signals to compare breathing coordination during maximal cycling in 12 people with asthma and 15 healthy controls. The study introduced BPratio, describing the balance between tidal volume and breathing frequency, and Ldis, a measure of thoraco-abdominal asynchrony. Most participants increased breathing depth before breathing frequency, while 33% of the asthma group accelerated breathing earlier in the workload. Deviant Ldis patterns were more common in asthma and were associated with reduced performance or greater post-exercise dyspnoea.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingHeart rate measured with Hexoskin was higher at midshift and end-shift when a preceding event had occurred. Cortisol also remained higher at those points, rather than showing the decline seen on shifts without an event. Heart rate and cortisol were summarized at defined time points; the observational analysis did not establish how long changes lasted after shifts, cumulative health effects or patient outcomes.

Summary

This planned secondary analysis followed 21 pediatric emergency-medicine providers over 91 eight-hour shifts. Hexoskin recorded heart rate under work clothes, while salivary cortisol was sampled separately at scheduled points. Shifts with a preceding event had higher heart rate and cortisol at midshift and end-shift than shifts without an event.

HexoskinArticleData Science & Artificial IntelligenceValidation & Methodology

Key FindingUsing waist acceleration from Hexoskin shirts worn by 13 healthy volunteers, a k-nearest-neighbor model classified activities with 96.37% accuracy. The researchers split samples at random, so the result does not show how well the model would classify the movements of a new participant.

Summary

Thirteen healthy volunteers each repeated 10 physical activities five times while wearing a Hexoskin garment with a waist-level accelerometer. The study classified these activities plus an unknown transition class using k-nearest-neighbor and support-vector-machine models. Overall test accuracy was 96.37% for the k-nearest-neighbor model and 95.4% for the support-vector-machine model under a random sample-level split.

HexoskinArticleRespiratory & Infectious DiseaseNeuroscience, Sleep & CognitionExercise Physiology & Sports ScienceClinical Medicine & Surgery

Key FindingAcross all 32 participants, fan-to-face reduced breathlessness by 0.5 Borg units versus no fan but not versus fan-to-leg; the exploratory subgroup with at least moderate breathlessness improved by 0.9 units versus no fan. Usable Hexoskin physiological data were available for 27 participants, and ventilation, breathing pattern, heart rate and blood oxygen saturation did not differ across conditions.

Summary

Thirty-two adults with COPD or interstitial lung disease completed randomized four-minute treadmill tests with fan-to-face, fan-to-leg and no-fan conditions. Two participants could not wear the Hexoskin shirt and three more had unusable software data, leaving usable Hexoskin physiological data for 27 participants. Fan-to-face reduced breathlessness by 0.5 Borg units versus no fan but did not outperform fan-to-leg; ventilation, breathing pattern, heart rate and blood oxygen saturation did not differ across conditions. Clinically meaningful reductions of at least one Borg unit occurred in 31% of the full sample and 45% of the more-breathless exploratory subgroup under fan-to-face versus no fan.

HexoskinArticleExercise Physiology & Sports ScienceValidation & Methodology

Key FindingReported agreement coefficients for Hexoskin heart rate versus the laboratory reference were 0.98–0.99 for almost all participants, but one participant had much lower values of 0.58 and 0.59 after sensor-contact problems. Breathing-rate coefficients ranged from 0.91 to 0.99, whereas estimated-ventilation coefficients ranged from 0.2 to 0.99 and remained difficult to validate. The healthy laboratory sample and progressive exercise protocol do not establish performance in clinical or daily-living settings.

Summary

The study compared Hexoskin heart rate, breathing rate and estimated ventilation with simultaneous laboratory references during a progressive maximal-exercise test. Results were reported for 28 healthy participants. Heart-rate and breathing-rate measurements generally showed high agreement, while estimated-ventilation results varied too widely for a firm validity conclusion.

HexoskinArticleOccupational Health & Human Factors

Key FindingThe exposure model, which combined particle measurements with Hexoskin inhalation rate, calculated the highest particle-size hazard levels for M20 and M25 concrete mixing, even though the highest particulate-matter emissions occurred during concrete-block cutting. Mixing also had higher inhalation rates than the other activities. These are modeled hazard indices from controlled dust-chamber tests, not observed disease rates or evidence that a control reduced exposure; field validation at construction sites remains needed.

Summary

In a controlled dust chamber, the study monitored particulate matter and inhalation rate across 15 construction activities and materials. Alphasense OPC-N3 sensors measured PM1, PM2.5 and PM10, while a Hexoskin wearable supplied inhalation-rate measurements. The authors used these measurements in a hazard-index model; the results do not represent disease incidence or demonstrate risk reduction at real construction sites.

HexoskinPreprintCardiology & ArrhythmiaRespiratory & Infectious DiseaseNeuroscience, Sleep & Cognition

HexoskinArticleOccupational Health & Human Factors

Key FindingIn a drilling study that paired particle sensors with Hexoskin inhalation-rate measurements, particulate-matter concentration was highest for M25 concrete blocks at 4.151 mg/m³ and lowest for solid blocks at 1.685 mg/m³. The paper reported aggregate toxic-substance hazard indices above its stated threshold across activities, while particulate-matter indices remained below it. These modeled classifications do not establish disease risk, regulatory compliance or performance at construction sites.

Summary

In controlled dust-chamber simulations, this study combined real-time inhalation-rate data with particulate-matter and toxic-substance measurements to model construction-activity health risks. Hexoskin measured inhalation rate, Alphasense OPC-N3 sensors measured particulate matter, and MiniVol sampling supported toxic-substance analysis. The resulting indices are modeled hazard estimates, not observed disease rates or evidence of reduced exposure.

HexoskinArticleCardiology & ArrhythmiaExercise Physiology & Sports ScienceData Science & Artificial Intelligence

Key FindingIn a comparison that included Hexoskin ECG, identifying activity before checking identity improved biometric verification relative to direct verification. The result concerns the analysis method, not a clinical diagnosis.

Summary

Researchers tested whether recognizing a person’s activity from an ECG before verifying identity improved biometric performance. They analyzed two public datasets, including ECG recorded with a Hexoskin Hx1 in the Vollmer dataset. Classifying activity first generally improved verification compared with using the ECG directly; a Naïve Bayes verifier generally produced lower equal-error rates than a decision tree.

HexoskinArticleData Science & Artificial IntelligenceOccupational Health & Human Factors

Key FindingThe stacked model for forecasting breathing rate from a Hexoskin dataset reported 8.1% on the study's SI error metric. One example trained on one minute of data and predicted the next four minutes. Support-vector-machine classification performed less well than other tested approaches; these results come from the study dataset and prototype and do not establish clinical accuracy, real-world hazard prevention or universal forecasting performance.

Summary

This study used a Hexoskin-derived dataset containing heart rate, breathing rate, minute ventilation, cadence and activity to build forecasting and action-state models. Random-forest and XGBoost models predicted heart and breathing rates up to 25 minutes ahead. A separate hardware prototype converted sustained predicted drops or spikes into haptic and sound alerts; the study did not demonstrate clinical or safety outcomes.

HexoskinArticleCardiology & ArrhythmiaExercise Physiology & Sports ScienceData Science & Artificial Intelligence

Key FindingScalogram images generally produced the lowest equal-error rates, while Mel-spectrograms produced the highest; Faros performed better with SGDM and Hexoskin with Adam. The best reported Hexoskin result was a 0.49% equal-error rate in the four-second scalogram standing condition, with 0.49% false acceptance and 0% false rejection across one, three and five genuine enrollment samples. Activity-aware verification generally outperformed direct verification, but performance varied widely across model settings.

Summary

This study investigated activity-aware electrocardiogram biometric verification using deep learning on wearable devices. It used ECG-only models in a framework designed to work with different device types, including a Hexoskin smart shirt with textile sensors. Activity-aware verification generally outperformed direct verification, although performance varied widely across image representations, window lengths and model settings.

HexoskinArticleData Science & Artificial IntelligenceOccupational Health & Human Factors

Key FindingIn a prototype simulation combining Hexoskin measurements with cameras and smart gloves, the system reported 85.4% accuracy for fatigue-signal detection. Reported errors for the capacitive heart-rate and breathing-rate estimates were 3.14 and 2.91, respectively; these were not accuracy measurements of the shirt. Unrealistic biometric changes and avatar motion limit the interpretation of the simulation.

Summary

Researchers built a digital twin of an industrial assembly task using physiological measurements from a Hexoskin vest, camera images and movement data from instrumented gloves. The prototype used these inputs to model movement and fatigue signals and trigger feedback. The authors observed unrealistic changes in simulated biometrics and avatar movement, limiting what the simulation can establish about workers.

HexoskinPreprintRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceData Science & Artificial Intelligence

HexoskinArticleClinical Medicine & SurgeryValidation & Methodology

Key FindingRoot mean square of successive differences showed a significant fixed bias between Hexoskin electrocardiogram-derived and RR-interval files, although Bland–Altman plots assessed several heart-rate-variability measures. Only heart rate and low-frequency power were significantly correlated between procedures. A discussion passage contradicts the latter result for low-frequency power, so these preliminary findings require cautious interpretation and standardized processing.

Summary

This preliminary study compared infant heart-rate and heart-rate-variability measures derived from electrocardiogram and RR-interval files after 5–10-minute resting recordings at one month of age. Researchers used a Hexoskin Shirt Junior, custom conversion scripts and Kubios software. Results differed for one variability measure, and only heart rate and low-frequency power were significantly correlated between file types.

HexoskinArticleClinical Medicine & SurgeryData Science & Artificial Intelligence

Key FindingUsing Hexoskin cardiac, respiratory, and activity inputs, the adaptive neuro-fuzzy model achieved 91% accuracy, 94% sensitivity, and 90% specificity for the reported prediabetes and type 2 diabetes detection task; Kappa testing reported 91% overall agreement with clinical assessment.

Summary

Baig and colleagues combined Hexoskin cardiac, respiratory, and activity measurements with blood glucose, demographics, clinical rules, and a knowledge base in an adaptive neuro-fuzzy model for early detection of prediabetes and type 2 diabetes. The approach was evaluated longitudinally in two people using 7.25 hours of real-time data and reported 91% accuracy, 94% sensitivity, 90% specificity, and 91% agreement with clinical assessment. The results support further evaluation of individualized wearable monitoring, while the two-person test limits generalizability.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingFor one surgeon performing total hip replacements, automated broaching took 3.7 minutes on average versus 6.1 minutes with manual broaching. Hexoskin estimates of the surgeon’s energy expenditure averaged 16.0 versus 32.6 Calories, respectively.

Summary

A single surgeon performed 26 direct-anterior total hip replacements with manual femoral broaching and 20 with a battery-powered automated system. Hexoskin-derived surgeon energy expenditure, heart rate and minute ventilation were compared during broaching. The automated method was associated with shorter broaching time and lower measured physiological demand in this observational, single-surgeon comparison.

HexoskinArticleClinical Medicine & SurgeryData Science & Artificial Intelligence

Key FindingA model combining Hexoskin measurements with clinical and other inputs achieved 91% overall agreement with expert assessment for early prediabetes and type 2 diabetes detection. In the two-person follow-up, the participant using the wearable and mobile-health approach had average HbA1c of 64 mmol/mol versus 87.5 mmol/mol in the participant receiving traditional management, but this descriptive comparison does not establish causality.

Summary

Baig and colleagues combined Hexoskin vital-sign and activity measurements with blood glucose, demographics, clinical guidelines, and a knowledge base in an adaptive neuro-fuzzy model for early detection of prediabetes and type 2 diabetes. The model was trained with data from 36 participants and tested longitudinally in two people. It reported 91% overall Kappa agreement and included a descriptive two-year follow-up.

HexoskinArticleValidation & Methodology

Key FindingHexoskin cadence increased in the expected order from rest to walking, running, and sprinting, with 95.5% of cadence variation explained by activity state. Breathing rate, minute ventilation, and accelerometer activity were significantly correlated with cadence, supporting their use for activity-pattern detection, whereas heart rate was the least consistent cadence correlate.

Summary

Banerjee and colleagues evaluated Hexoskin as a prospective unobtrusive monitoring platform for dementia care, first validating cadence in four healthy adults across rest, walking, running, and sprinting. Cadence changed in the expected order across the four activity states. This preliminary validation did not test people with dementia.

HexoskinArticleOccupational Health & Human FactorsData Science & Artificial IntelligenceValidation & Methodology

Key FindingOperator engagement was higher in manual work than in the automated condition on the Utrecht Work Engagement Scale absorption subscore, 4.21 versus 3.32. The logistic-regression model using Hexoskin respiration and acceleration predicted engagement states with 80.95% accuracy under leave-one-out cross-validation. The feedback display is a prototype based on respiration and acceleration; the study does not establish improved engagement, productivity or safety.

Summary

This study developed a logistic-regression model to estimate manufacturing-operator engagement from respiration and acceleration data collected with Hexoskin. Under leave-one-out cross-validation, the model reported 80.95% accuracy. The prototype displayed engagement on a continuous color gradient calibrated to the range observed in its training data.

HexoskinArticleCardiology & ArrhythmiaRespiratory & Infectious DiseaseClinical Medicine & SurgeryValidation & Methodology

Key FindingIn 31 older adults, processed Hexoskin cadence, respiration rate and heart rate tracked their corresponding reference measures, while minute ventilation and short-duration activity level did not. Heart rate and processed cadence showed the strongest agreement, and respiration rate showed a moderate but significant association with low bias. The results support selective use of the validated signals and do not establish that every commercial-sensor output is clinically interchangeable with its reference.

Summary

Banerjee and colleagues evaluated Hexoskin against laboratory reference measures in 31 adults aged 65 years or older, including participants with mobility impairment, cognitive impairment, both, or neither. Processed cadence, respiration rate, and heart rate showed useful agreement with their references, while minute ventilation and short-duration activity level did not. The findings support selective longitudinal monitoring in older adults and emphasize signal-specific validation before dementia-management use.

HexoskinArticleOccupational Health & Human FactorsData Science & Artificial Intelligence

Key FindingThe Optimal State Deviation Index, developed using Hexoskin data during manual and automated snowshoe assembly, achieved 81.31% accuracy on training data and 80.95% under leave-one-out testing. The formula used the full Task 1 and Task 2 dataset, so the test estimate comes from cross-validation rather than a separate external cohort. In display testing, participants found the discrete color gradient more distracting; the study did not demonstrate improved workplace performance or safety.

Summary

Twenty-two participants completed manual and fully automated snowshoe-assembly tasks while wearing Hexoskin for physiological data collection. Those data were used to build the Optimal State Deviation Index for estimating engagement. The model reported 80.95% test accuracy under leave-one-out cross-validation. A separate prototype displayed engagement through a color gradient as part of an iterative biocybernetic-feedback design.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingIn five surgeons monitored with Hexoskin, direct-anterior procedures had a mean stress index of 17.4 versus 12.4 for posterior procedures. Heart rate averaged 101 versus 98.3 beats/min, and minute ventilation 21.7 versus 18.7 L/min. Operative time was shorter for the direct-anterior approach; results came from five high-volume surgeons and may not generalize to other surgeons, settings or patient outcomes.

Summary

Five high-volume surgeons were monitored with Hexoskin during 144 consecutive total hip arthroplasties: 67 direct-anterior and 77 posterior procedures. Direct-anterior cases were shorter but were associated with higher surgeon stress index, heart rate and minute ventilation. This observational comparison concerns surgeon physiology and does not establish patient outcomes or that surgical approach caused the differences.

HexoskinArticleExercise Physiology & Sports ScienceOccupational Health & Human Factors

Key FindingHexoskin accelerometry showed higher average, RMS, and RM4 acceleration during training than shows, while artists rated shows as more exerting and time within the five acceleration ranges did not differ significantly. These track-level profiles suggest wearable inertial data can help coaches characterize mechanical demands in theatre environments where GPS is unavailable.

Summary

Barker and colleagues used Hexoskin's hip-mounted tri-axial accelerometer to compare seven male acrobats' trampoline performances across training and shows. Training produced higher average, RMS, and RM4 acceleration, whereas perceived exertion was higher during shows and time in defined acceleration ranges did not differ significantly. The study positions wearable inertial measurement as a practical workload-management tool for theatre and arena performances where GPS is inoperable.

HexoskinArticleExercise Physiology & Sports ScienceValidation & Methodology

Key FindingBody heat storage was similar with Hexoskin alone and the semi-nude control, 214 versus 211 kJ. Under a work uniform, storage was also similar with Hexoskin and a cotton undershirt, 307 versus 318 kJ. Evaporative, dry and total heat loss did not differ significantly between the paired conditions; this controlled study in nine men does not establish results for other populations, clothing systems or environments.

Summary

Nine healthy young men completed four randomly ordered exercise sessions in hot, dry conditions. Each session included seated rest, 60 minutes of semi-recumbent cycling at a fixed metabolic heat production and 60 minutes of recovery. Comparisons found no evidence that Hexoskin, worn alone or under a work uniform, impaired whole-body heat dissipation under the tested conditions.

HexoskinConference AbstractOccupational Health & Human FactorsClinical Medicine & Surgery

HexoskinAcademic ThesisRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceValidation & Methodology

Key FindingHexoskin and Vmax showed similar heart-rate and breathing-rate responses during exercise in the COPD cohort. The small rehabilitation subgroup and the less consistent lung-volume results limit conclusions about using the devices interchangeably.

Summary

This master’s thesis compared Hexoskin with a Vmax metabolic cart during incremental and constant-workload cycling in adults with chronic obstructive pulmonary disease. Fourteen participants wore the shirt; seven contributed to the assessment of change after pulmonary rehabilitation. Heart rate and breathing rate changed with exercise, without significant device or device-by-time effects across the three tests. The findings for tidal volume and inspiratory capacity were less consistent.

HexoskinArticleExercise Physiology & Sports ScienceRespiratory & Infectious Disease

Key FindingHeart rate measured with Hexoskin increased similarly across the N95, cloth-mask and no-covering conditions, with no significant difference in slopes. Breathing rate rose in all conditions, but the increase was steeper without a covering than with a cloth mask. Transcutaneous carbon dioxide decreased over time only in the no-covering condition; results from 12 athletes and a maximal treadmill protocol do not establish effects in other populations or everyday activity.

Summary

Twelve Division I collegiate athletes completed graded treadmill tests while wearing an N95 respirator, a cloth mask or no face covering. Hexoskin continuously measured heart rate and breathing rate. Exercise lasted longer without a face covering; heart-rate slopes were similar across conditions, while breathing rate increased more steeply without a covering than with a cloth mask.

HexoskinArticleExercise Physiology & Sports ScienceData Science & Artificial Intelligence

Key FindingAerobic-system mean normalized gain derived from predicted oxygen uptake during unsupervised daily activity correlated with the laboratory-derived measure at r=0.82 (p<0.001; n=13), with a consistent difference of about 16%. The model distinguished light, moderate and vigorous activity, but only about 20% of monitored data occurred while participants were active enough for reliable aerobic-system analysis. Hexoskin supplied continuous activity, breathing and heart-rate inputs, while free-living oxygen uptake was model-predicted and the proposed health-status application remained speculative.

Summary

Beltrame and colleagues monitored 13 healthy active men with Hexoskin for about six hours per day over four days to study aerobic-system dynamics during unsupervised daily activity. Hexoskin cardiac, respiratory and movement features were used to predict free-living oxygen uptake and derive aerobic-system dynamics. The predicted daily-life temporal measure correlated with the directly measured laboratory measure at r=0.82 (p<0.001).

HexoskinArticleOccupational Health & Human Factors

Key FindingIn 32 paramedics monitored with Hexoskin, posting, body mass index and call type were associated with physiological responses during shifts and emergency calls. Greater cardiorespiratory fitness was associated with lower physiological demand among the 14 participants who completed fitness testing. These observational findings do not show that posting, body mass or fitness caused the measured differences.

Summary

A field study monitored 32 paramedics with Hexoskin around work shifts and emergency calls, recording heart rate, respiratory rate and occupational acceleration. Fourteen also completed maximal cardiorespiratory fitness testing. The study examined how posting, body mass, call characteristics and fitness were associated with physiological responses during paramedic work.

HexoskinArticleExercise Physiology & Sports ScienceData Science & Artificial Intelligence

Key FindingIn 16 healthy active men, the Hexoskin-feature random-forest model predicted oxygen uptake during simulated daily activities with r=0.87 versus COSMED measurements and produced mean normalized gain correlated at r=0.71 with the measured endpoint. The model distinguished resting, light, and moderate metabolic demands, but intense activity represented less than 2% of the data and was excluded from the validated range.

Summary

Beltrame and colleagues evaluated oxygen-uptake prediction in 16 healthy active men during controlled random-paced walking and approximately 20 minutes of simulated daily activities. Hexoskin cardiac, respiratory, and motion features supplied the wearable inputs, while a COSMED K4b2 system provided the directly measured oxygen-uptake reference. The findings support prediction under the tested activity conditions and identify the need for further reliability testing before broader application.

HexoskinArticleOccupational Health & Human Factors

Key FindingTen of 16 participants increased oxygen uptake in the active chair, two showed no change and four showed a decrease; the average change was 0.02 ± 0.04 L/min. Hexoskin cardiac measurements showed larger relative heart-rate-variability decreases in males than females, although absolute between-sex differences were not significant. This short laboratory study does not establish long-term health, productivity or weight-management benefits.

Summary

Sixteen participants completed 20-minute standardized data-entry tasks in an active chair and a conventional chair. A metabolic cart measured oxygen uptake and ventilation, while Hexoskin measured heart rate and heart-rate variability. On average, active sitting increased oxygen uptake by 0.02 L/min and energy expenditure by 1.5 kcal per 15 minutes, with substantial individual variation.

HexoskinArticleRespiratory & Infectious DiseaseValidation & Methodology

Key FindingIn a 10-minute seated test, 94.5% of breathing-rate estimates from a 77 GHz radar and nonlinear least-squares model were within one breath per minute of the Hexoskin reference. Across the tested scenarios, the authors reported accurate estimation and improvement over discrete Fourier transform and frequency-time phase-regression methods, but the laboratory evidence does not establish performance in a broader population or free-living setting.

Summary

This study used Hexoskin as a reference sensor to evaluate non-contact breathing-rate estimation with a frequency-modulated continuous-wave radar. A nonlinear least-squares estimator was compared with discrete Fourier transform and frequency-time phase-regression approaches in static, texting, and typing conditions. The proposed method produced high within-one-breath-per-minute accuracy in the laboratory tests.

AstroskinPoster / PresentationOccupational Health & Human FactorsValidation & MethodologyExercise Physiology & Sports Science

Key FindingThe team obtained usable physiological data across 24-hour wake and sleep periods, and crew described the prototype as comfortable and easy to use. The study tested feasibility in a ground analog.

Summary

Eight crew members wore an Astroskin prototype during two 30-day missions in NASA’s ground-based Human Exploration Research Analog. The team assessed wake and sleep recordings, exercise monitoring, and how easily crew could use the system.

HexoskinArticleCardiology & ArrhythmiaValidation & Methodology

Key FindingThe narrow-band thermal signal and Hexoskin ECG both oscillated at 0.9 Hz, approximately 54 beats/min, and the authors reported a strong correlation. This was a seated single-subject proof of concept.

Summary

Bennett and colleagues evaluated whether thermal video could recover heart rate using adaptive Eulerian Video Magnification. A seated subject wore a Hexoskin smart shirt to provide the ECG reference while light and thermal cameras recorded simultaneously. Wide-band processing identified a chest region of interest, and narrow-band thermal processing produced an oscillation at 0.9 Hz, matching the ECG frequency of about 54 beats/min. The proof of concept supports further study of contactless physiological-signal extraction but does not establish validation across people or conditions.

HexoskinAcademic ThesisRespiratory & Infectious DiseaseValidation & Methodology

Key FindingNine of 12 Hexoskin shirts showed a linear response to changes in torso circumference on the simulator. Temperature correction reduced median drift from −34.7 to −0.5 mL/°C, identifying a source of measurement error that can be tested independently of breathing changes.

Summary

This doctoral thesis examined whether Hexoskin respiratory measurements were stable enough for monitoring changes in lung volume. Researchers used a torso simulator to test the shirts’ response to changes in temperature and circumference. A separate experiment compared shirt measurements with spirometry during seven everyday tasks in 15 healthy adults. The human experiment did not establish accuracy in patients with COPD.

HexoskinArticleCardiology & ArrhythmiaExercise Physiology & Sports Science

Key FindingHexoskin HRV showed a moderate association with esports activity and a team-level correlation of r=0.78 between winning and losing teams' global HRV patterns. The winning team had lower mean SDRR, but substantial individual variability and absent high-resolution gameplay context prevented a confident link between HRV and performance.

Summary

Bennett and colleagues used Hexoskin to monitor ECG-derived beat intervals and heart-rate variability during 1.5- to 2.5-hour League of Legends and Overwatch sessions. The results describe 19 male players across two cohorts, while the limitations describe n=10 and two device failures, leaving an unresolved internal count discrepancy. Winning-team mean SDRR was lower, but substantial individual variability and missing high-resolution gameplay context prevented a confident link between heart-rate variability and performance.

HexoskinConference AbstractExercise Physiology & Sports Science

HexoskinArticleCardiology & ArrhythmiaRespiratory & Infectious DiseaseNeuroscience, Sleep & CognitionOccupational Health & Human Factors

Key FindingThe narrative review treats challenge and threat as physiologically distinguishable stress appraisals and surveys cardiovascular and hormonal measurements that could be combined in high-stress paramedicine scenarios. Hexoskin is presented as one portable cardiovascular-monitoring option, not as a device validated by a new trial in this paper.

Summary

Betson and colleagues narratively reviewed approaches for quantifying threat-or-challenge responses among undergraduate paramedicine students during high-stress clinical scenarios. The review is motivated by the mental, physical, and emotional demands of paramedicine practice. Within its survey of cardiovascular measurement options, the review identifies Hexoskin as a portable technology for monitoring cardiovascular responses.

HexoskinArticleExercise Physiology & Sports Science

Key FindingMean heart rate measured with Hexoskin was nearly 14% higher and peak heart rate nearly 19% higher during pickleball doubles than walking. Estimated caloric expenditure was 36% higher during pickleball, while walking produced 54% more steps. The calorie values were movement-based predictions rather than oxygen-consumption measurements, and results from 25 healthy adults do not establish long-term health or weight-management benefits.

Summary

Twenty-five healthy adults completed 30-minute pickleball-doubles and self-paced walking sessions on the same day in counterbalanced order. Hexoskin provided heart rate, step count and movement-based caloric-expenditure estimates. Pickleball produced higher heart rate, estimated caloric expenditure, perceived exertion and enjoyment, while walking produced more steps.

HexoskinArticleCardiology & ArrhythmiaRespiratory & Infectious DiseaseNeuroscience, Sleep & CognitionClinical Medicine & Surgery

Key FindingTwelve weeks of slow breathing reduced psychological stress within both groups, but extending exhalation did not provide a significant incremental benefit over equal-ratio breathing. Neither intervention significantly changed high-frequency heart-rate variability, while Hexoskin confirmed adherence to the assigned home breathing patterns.

Summary

Birdee and colleagues conducted a 12-week randomized, single-blinded trial comparing extended-exhale and equal-ratio slow breathing. The abstract reports 100 participants, while the article body reports 99 enrolled, 97 in the intention-to-treat analysis and 94 with complete data. Participants wore Hexoskin shirts during home practice to assess adherence to the assigned breathing ratios. Slow breathing reduced PROMIS Anxiety within both groups, while neither strategy significantly changed heart-rate variability and the between-group differences were nonsignificant.

HexoskinArticleOccupational Health & Human FactorsValidation & Methodology

Key FindingThe Hexoskin garment was reported as comfortable and did not impede the wearer’s field-training tasks; no apparent failure or breakdown was observed during the exercise. It was worn over bare skin beneath a cotton T-shirt, fatigue top, rain jacket and tactical load-bearing vest with blank ammunition, without body armour or ballistic plates. This single-person training report does not establish performance across personnel, protective configurations or operational deployments.

Summary

One Canadian military section commander wore Hexoskin during a reserve-unit field exercise in British Columbia. The report describes 30 hours of continuous recording and a Bluetooth field test during a simulated mass-casualty event. The wearer reported comfort, no interference with duties and no apparent equipment failure during the exercise.

HexoskinArticleCardiology & ArrhythmiaNeuroscience, Sleep & CognitionExercise Physiology & Sports Science

Key FindingIn a walking study that used Hexoskin to track heart rate during the tasks, higher heart-rate reserve was associated with greater prefrontal-cortex activation and shorter stride time during the more difficult dual-task walking conditions. Greater HbO2 activation was also associated with higher MoCA and Mini-BEST scores and faster walking, but cognitive error-rate effects were not significant.

Summary

Bishnoi and Hernandez examined how heart-rate reserve was associated with prefrontal-cortex activation and gait during increasingly difficult Trail Walking Tasks. Heart-rate reserve was calculated from baseline resting heart rate and a formula-based maximum, while Hexoskin recorded heart-rate responses during the walking tasks and fNIRS measured brain activation. Higher heart-rate reserve was associated with greater prefrontal activation and shorter stride time, while stride width and stride length showed no significant heart-rate-reserve or task effects.

HexoskinConference AbstractNeuroscience, Sleep & Cognition

HexoskinArticleCardiology & ArrhythmiaRespiratory & Infectious DiseaseNeuroscience, Sleep & CognitionExercise Physiology & Sports ScienceOccupational Health & Human Factors

Key FindingIn a study using Hexoskin for heart and respiratory rates and separate sensors for brain oxygenation, higher immersion was associated with increased oxygenation in three of four prefrontal-cortex regions and an increase in heart rate from 86.70 to 87.66 beats per minute. Respiratory rate decreased from 18.49 to 17.84 breaths per minute, a statistically significant but practically negligible change contrary to the hypothesis; bilateral Index of Cognitive Activity also changed in the opposite direction from expected. The two self-report scales showed no significant difference, and the nine-participant findings require further validation.

Summary

Nine law-enforcement trainees aged 21 to 31 completed one-screen and five-screen use-of-force simulator conditions. Multimodal sensors tracked cognitive-workload biomarkers, with Hexoskin supplying heart and respiratory rates. Under higher immersion, prefrontal oxygenation and heart rate increased, while respiratory rate decreased. The findings support feasibility in this preliminary setting but require confirmation in a larger sample.

HexoskinArticleNeuroscience, Sleep & CognitionData Science & Artificial Intelligence

Key FindingIn 78 hospitalized people with epilepsy, forecasts using Hexoskin sleep measurements beat chance for approximately 49% of patients at 16 hours and 40% at 24 hours. These percentages are the shares of patients whose forecasts outperformed chance, not the proportions of seizures predicted.

Summary

The study used Hexoskin nocturnal sleep features to forecast seizure days in 78 hospitalized people with epilepsy. Support-vector-machine models normalized features to reference nights and evaluated 16-hour and 24-hour horizons with nested leave-one-patient-out cross-validation. The authors reported improvement over chance for some patients and called for longer recordings in residential settings.

HexoskinPoster / PresentationCardiology & ArrhythmiaRespiratory & Infectious DiseaseOccupational Health & Human Factors

HexoskinAcademic ThesisExercise Physiology & Sports ScienceNeuroscience, Sleep & Cognition

Key FindingSome participants reported a mismatch between their breathing and the Path triangle, while others deliberately varied their breathing to explore the Trees response. Arrow cues helped several runners understand inhale-and-exhale timing or find a more intuitive rhythm. These qualitative, self-reported experiences from a laboratory session do not establish improved breathing, running performance or training outcomes.

Summary

Runners used real-time Hexoskin breathing feedback during treadmill running through three visual displays: ambient Trees, directional Arrows and a continuous Path. Speeds were adjusted individually in a sport-science laboratory. Interviews, body maps and questionnaires described different ways participants perceived, adapted to and interacted with the breathing visualizations.

HexoskinArticleCardiology & ArrhythmiaNeuroscience, Sleep & CognitionExercise Physiology & Sports ScienceOccupational Health & Human FactorsValidation & Methodology

Key FindingHexoskin provided acceptable ECG performance overall, with 91.33% sufficient morphological signal quality and a QRS positive predictive value of 0.989. Sufficient signal quality was 99.61% at rest, 86.82% during level walking, 99.15% during the two-back task and 77.60% uphill. Movement produced more noise, and close shirt fit mattered because electrode displacement could reduce accuracy; Faros produced the most stable results in this simultaneous comparison.

Summary

Bläsing and colleagues simultaneously compared Hexoskin Hx1 with four other wearable ECG devices in 13 adults across four five-minute phases: rest, walking, cognitive workload and uphill walking. They assessed local ECG noise, QRS detection and RR-interval quality with morphSQ and a modified Smith-Waterman approach. Hexoskin achieved 91.33% sufficient morphological signal quality overall and a QRS positive predictive value of 0.989. Signal quality was highest at rest and during the standing two-back task and lower during level and uphill walking, showing that performance depended on the activity.

HexoskinArticleRespiratory & Infectious DiseaseData Science & Artificial IntelligenceValidation & Methodology

Key FindingThe cough-detection model using Hexoskin acceleration and respiration reported an F1 score of 93.0 ± 1.1%, compared with 93.1 ± 1.3% when all sensors were used. Acceleration- and respiration-based models outperformed electrocardiogram-based models for this task. The study used prompted tasks in 44 healthy adults under stationary conditions; it did not establish performance during substantial movement or in patients with spontaneous cough.

Summary

In an observational study, 44 healthy adults wore a Hexoskin garment while performing prompted coughs and other respiratory tasks in supine and seated positions. A random-forest model using the garment’s acceleration and respiratory-inductance features reported an F1 score of 93.0 ± 1.1%, precision of 84.2 ± 4.2% and sensitivity of 95.5 ± 1.6%. Patient monitoring was proposed as future work.

HexoskinDatasetCardiology & ArrhythmiaNeuroscience, Sleep & CognitionValidation & Methodology

HexoskinArticleCardiology & ArrhythmiaData Science & Artificial Intelligence

Key FindingA classifier using Hexoskin ECG recordings from 38 participants achieved an area under the ROC curve of 0.86 for distinguishing faster from slower heart-rate recovery after exercise. The cutoff of 28 beats/min was the median in this group, not a validated clinical risk threshold.

Summary

The study used electrocardiogram signals recorded by sensors embedded in a Hexoskin shirt to classify heart-rate recovery in 38 participants after exercise. Groups were defined using the cohort median recovery value of 28 beats/min, and a support-vector classifier reported an area under the curve of 86%. The cohort-relative labels are not a clinical diagnosis or externally validated risk threshold.

HexoskinAcademic ThesisExercise Physiology & Sports ScienceClinical Medicine & Surgery

Key FindingAll eight initial video segments had a camera-derived frequency peak matching the manually annotated truth peak, and sparse optical flow produced the lowest sum of squared errors among the tested computer-vision methods. Hexoskin accelerometer Y- and Z-channel features each correlated 0.954 with treadmill-derived gait speed at the 0.01 significance level, while the camera V-channel feature was not significantly correlated with true speed. Generated and reference waveforms had coherence of 1 at the gait-frequency peak for each participant, and the author concluded that either sensor alone or both together could be considered under the controlled conditions.

Summary

Schneider evaluated gait analysis from a wearable camera embedded in glasses and a Hexoskin accelerometer vest. Four participants supplied the initial video-only validation data, and five participants completed an additional treadmill collection at five speeds from 2.0 to 3.5 mph with both sensors. Computer-vision and signal-processing methods recovered walking pace in the controlled treadmill setting, and Hexoskin acceleration features were related to treadmill-derived gait speed. The author reported that one sensor could support parametric gait analysis in this setting, while retaining both could provide richer data.

HexoskinArticleOccupational Health & Human FactorsValidation & Methodology

Key FindingIn a field study using Hexoskin ECG, higher-stress phases were observed, but the researchers could not identify them clearly for three of the five nurses. Low activity was also an unreliable marker of breaks because it occurred during documentation, telephone calls and medication preparation. The rrHRV method was described as relatively robust to movement artifacts because it calculated values only when two adjacent R peaks were detected.

Summary

This nursing field study used Hexoskin shirts to record raw electrocardiogram data for heart-rate-variability analysis during morning shifts and assessed whether the measures could help describe stressful events. The analysis also considered physical activity and the ambiguity of low-strain periods. Its findings caution against treating heart-rate-variability patterns as a direct label for breaks or stress.

HexoskinAcademic ThesisRespiratory & Infectious DiseaseData Science & Artificial IntelligenceValidation & Methodology

Key FindingCoherence decreased under total obstruction in some comparisons, while overall gain and phase-angle effects were inconsistent across posture and obstruction conditions. Low coherence produced missing phase and gain values, and participant responses varied substantially. The weighted vest reproduced only added thoracic and abdominal load, and the four-person healthy sample does not establish clinical detection of paradoxical breathing or obesity-related effects.

Summary

This exploratory study used Hexoskin respiratory-inductance signals from four healthy women to examine time-frequency measures of paradoxical breathing during sitting and lying simulations with different airway resistances and a weighted vest. Outcomes included phase angle, gain and coherence. The small awake simulation was not a diagnostic study in patients with sleep apnea, asthma or obesity.

HexoskinArticleCardiology & ArrhythmiaData Science & Artificial IntelligenceValidation & Methodology

Key FindingOn single-lead Hexoskin ECG, xqrs had the highest combined precision and sensitivity among the 12 detectors, with mean sensitivity 0.996 and precision 0.991 in the primary condition. For three participants with more false positives, RR-interval editing improved precision but reduced sensitivity. False-negative interpolation remained future work, and the 13-person healthy dataset does not establish performance in post-hospital patients.

Summary

Researchers benchmarked 12 open-source QRS-detection algorithms and a downstream RR-interval editing step using single-lead Hexoskin ECG from 13 healthy adults performing several activities. The comparison used a 100 ms matching tolerance and included added baseline, muscle and electrode-motion noise. It was an algorithm benchmark rather than a clinical diagnostic or patient-outcome study.

HexoskinArticleExercise Physiology & Sports ScienceValidation & Methodology

Key FindingIn 10 elite male cyclists, Hexoskin heart rate correlated at least 0.98 with the laboratory reference across work rates. Breathing frequency correlated at least 0.95 through 75% of maximum work rate, but estimated minute ventilation had larger errors, particularly at high intensity.

Summary

Ten male elite cyclists completed two progressive cycling tests to exhaustion 3–4 days apart. Hexoskin heart rate, breathing frequency and estimated minute ventilation were compared with laboratory measurements. Heart rate and breathing frequency showed strong agreement through most work-rate stages, while ventilation estimates were less accurate, particularly at high intensity.

HexoskinAcademic ThesisClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingElevated anxiety was reported in 49 of 56 assessment participants, and 33% did not pass the final-term assessment. Heart rate was higher in the assessment group before and during the activity, but not afterward; heart-rate-variability findings were mixed. The mock-assessment control could also experience stress, and the observational design does not show that anxiety caused performance differences.

Summary

This exploratory study compared 56 undergraduate paramedic students completing final-term clinical assessments with 15 students completing mock classroom assessments. Hexoskin recorded heart rate, respiratory rate and heart-rate variability; questionnaires and saliva assays supplied separate psychological and biomarker measures. Valid Hexoskin streams were available for 53 participants. The groups were purposefully sampled rather than randomly assigned.

HexoskinConference AbstractClinical Medicine & SurgeryOccupational Health & Human Factors

HexoskinAcademic ThesisExercise Physiology & Sports ScienceData Science & Artificial IntelligenceValidation & Methodology

Key FindingAcross three exercise conditions, predicted oxygen uptake had a repeated-measures correlation of 0.911 with direct measurements and bias of 0.022 L/min. Predicted response kinetics were slower and only moderately correlated with measurements. The model required individual maximum heart rate and calibrated ventilation, limiting use outside the laboratory, and the distinct study and tuning groups must not be combined.

Summary

This thesis included two related exercise studies. One used eight healthy adults to evaluate a frequency-domain indicator of oxygen-uptake response kinetics; the other used 15 healthy adults to develop a random-forest oxygen-uptake model from Hexoskin-derived heart-rate and respiratory features. A separate seven-person dataset selected the tree count, and model performance was evaluated with leave-one-participant-out testing.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingThe protocol specifies Hexoskin cardiac monitoring and would enroll 42 physicians per group and compare a three-hour CRI:SIS intervention with control using on-shift heart-rate variability and post-shift anxiety. Four shift recordings and five-minute analysis periods were specified. These are planned methods, not completed enrollment, measurements or treatment effects; the protocol does not establish reduced stress, improved preparedness or patient benefit.

Summary

This study protocol combines interviews and field observation to refine a simulation-preparedness program for emergency physicians, followed by a planned randomized trial. Participants would complete four shift recordings, with randomization after the first two. Hexoskin would measure heart rate and heart-rate variability, while post-shift anxiety would be measured separately with the State-Trait Anxiety Inventory.

HexoskinAcademic ThesisExercise Physiology & Sports ScienceOccupational Health & Human Factors

Key FindingAll eight participants completed the exercise bouts with upper-body and whole-body cooling, while none began the third no-cooling bout. Both cooling systems reduced perceived exertion and breathing discomfort relative to no cooling, but skin-wetness responses differed: wetness rose during the first bout with no cooling and upper-body cooling, but not with whole-body cooling. The additional whole-body-cooling advantage for limiting core-temperature rise appeared only in the third bout, and applicability to operational firefighting tasks remains uncertain.

Summary

Eight healthy adults completed exercise in heat while wearing firefighter turnout gear under no-cooling, upper-body-cooling and whole-body-cooling conditions. A Hexoskin tank-top measured heart rate, while separate instruments measured core temperature, skin temperature and oxygen consumption. All participants completed three exercise bouts with both cooling systems, whereas none began the third no-cooling bout. Both cooling conditions reduced several measures of strain relative to no cooling, but whole-body cooling showed an additional core-temperature advantage only during the third bout.

HexoskinArticleClinical Medicine & SurgeryCardiology & Arrhythmia

Key FindingHeart-rate-variability measures derived from Hexoskin cardiac recordings showed mixed results. RMSSD decreased across all 64 participants from the initial to final measurement, and some frequency-domain differences appeared between manipulation and simulated-manipulation conditions. The study did not provide consistent evidence that the techniques changed heart-rate variability; the healthy 25–35-year-old sample and different assignment process for the no-intervention group limit conclusions about treatment effects.

Summary

This pilot study examined heart-rate variability after visceral osteopathic manipulations directed at pericardial attachments. Sixty-four healthy adults aged 25–35 were assigned to manipulation, simulated-manipulation or no-intervention groups. Hexoskin recorded heart rate, and five-minute segments were analyzed in time and frequency domains over several study visits.

HexoskinConference AbstractExercise Physiology & Sports ScienceOccupational Health & Human Factors

HexoskinConference AbstractClinical Medicine & SurgeryRespiratory & Infectious DiseaseExercise Physiology & Sports Science

HexoskinArticleExercise Physiology & Sports ScienceData Science & Artificial IntelligenceValidation & Methodology

Key FindingEstimates of VO2max based on longitudinal Hexoskin data and demographic measures correlated with measured VO2max (Pearson r = 0.804), with mean absolute error of 0.38 ± 0.13 L/min and mean difference of 0.038 L/min. Age ranked highest among the model inputs, followed by heart rate, height, weight and acceleration. Internal cross-validation does not establish external performance or accuracy within the study's small clinical subgroups.

Summary

The study combined seven days of longitudinal Hexoskin heart-rate, respiratory and movement data with demographic measures from 43 participants. These inputs were used to estimate cardiopulmonary-exercise-test VO2max. A support-vector-regression model was evaluated with nine-fold cross-validation, and SHAP analysis described the contribution of model inputs.

HexoskinStudy ProtocolRespiratory & Infectious DiseaseValidation & Methodology

HexoskinArticleNeuroscience, Sleep & Cognition

Key FindingLower sleep efficiency measured by Hexoskin was associated with greater odds of a seizure within 24 hours after awakening; higher wake time after sleep onset showed a similar association. Nights containing a nocturnal seizure also had lower sleep efficiency, more wake time and higher odds of a subsequent seizure. These inpatient night-level associations do not establish causality or prospective forecasting performance at home.

Summary

The study analyzed 304 nights from 47 people with confirmed epilepsy admitted to an inpatient monitoring unit. Hexoskin sleep measures were compared with seizures annotated from video EEG. Six patients with seizures every day or night were excluded, and nights were classified by whether a seizure occurred within 24 hours after awakening.

HexoskinArticleCardiology & ArrhythmiaData Science & Artificial IntelligenceValidation & Methodology

Key FindingOn the public databases, reported maximum F-scores reached 94.6% for AFPDB and 99.8% for AFTDB under selected settings. These were repeated cross-validation results from database signals, not Hexoskin performance measures or an independent challenge test. The garment example showed normal-rhythm patterns only and did not establish atrial-fibrillation detection accuracy or clinical validation.

Summary

The study proposed a dynamic symbolic assignment method that converts ECG RR intervals into transition features for classifying normal sinus rhythm and paroxysmal atrial fibrillation. Machine-learning performance was evaluated with public PhysioNet databases. Hexoskin and a capacitive ECG chair were used separately to demonstrate normal-rhythm transition patterns.

HexoskinAcademic ThesisData Science & Artificial IntelligenceOccupational Health & Human Factors

Using Deep Learning for Motion Analysis of 3D Motion Capture Data for Forecasting Motion and Fatigue

Key FindingThe preferred motion-forecasting settings used a 100 ms horizon for LSTM and a 50 ms horizon for GRU. The fatigue models did not achieve the expected result: their predictions settled near a Borg rating of 14 and did not track higher reported exertion. Findings from two test participants in a university sample do not establish performance for experienced material-handling workers or injury prevention.

Summary

This thesis used motion-capture data from 10 university participants to test LSTM and GRU models for short-term lifting-motion forecasting and prediction of self-reported exertion. Eight participants supplied training data and two supplied test data. Hexoskin heart-rate and breathing-rate measurements were collected but explicitly excluded from the analysis.

HexoskinArticleOccupational Health & Human FactorsData Science & Artificial IntelligenceCardiology & ArrhythmiaValidation & Methodology

Key FindingIn a comparison using features derived from Hexoskin heart-rate data, the random forest produced the highest reported correct-entry rate at 86.3%, while CatBoost reached 85.3% and processed each record in about 2.02 ms. More than 70% of segments represented low stress ratings of 2–3, limiting the range used for training and evaluation. The small study does not establish clinical stress diagnosis or performance across more diverse stress levels.

Summary

Nineteen healthy adults aged 19–55 contributed laboratory or daily-life recordings for a wearable stress-prediction study. Fifteen completed laboratory testing, and eight contributed field recordings. Hexoskin heart-rate data were converted into heart-rate-variability features and paired with self-reported stress ratings for comparison of five regression models.

HexoskinArticleCardiology & ArrhythmiaRespiratory & Infectious DiseaseOccupational Health & Human FactorsValidation & Methodology

Key FindingDuring dry immersion, heart rate measured with Hexoskin fell by a further 5–10 beats/min, while SDNN and sample entropy changed significantly. Respiratory rate decreased from 16 to 13 breaths/min, but this change was not statistically significant, and several other heart-rate-variability measures did not change significantly. The small healthy cohort and ground-based analogue do not establish clinical or spaceflight performance.

Summary

Eight healthy young adults wore a Hexoskin Smart Shirt during a 45-minute dry-immersion session. The session was used as a ground-based model of weightlessness. ECG-derived heart-rate variability and respiratory rate were analyzed in five-minute segments, with motion during blood-pressure measurement introducing some signal noise.

HexoskinArticleOccupational Health & Human FactorsData Science & Artificial IntelligenceCardiology & ArrhythmiaValidation & Methodology

Key FindingThe HRVEST workflow for Hexoskin cardiac recordings processed 413 ECG recordings in an average of 15 minutes 28 seconds on the reported laptop and retained 99.7% usable data after artifact correction. Run time increased roughly linearly with file count, but depended on hardware and selected options. This was a processing-methods evaluation, not validation of physiological stress measurement or a clinical outcome study.

Summary

Researchers developed HRVEST to process Hexoskin ECG from a clinical trial involving 81 emergency physicians and 324 work-shift recordings. The open-source workflow converted files and synchronized metadata. It then divided recordings into five-minute windows, corrected artifacts, classified signal quality and calculated heart-rate-variability measures from a 347 GB dataset.

HexoskinConference AbstractNeuroscience, Sleep & CognitionData Science & Artificial IntelligenceValidation & Methodology

HexoskinArticleNeuroscience, Sleep & CognitionData Science & Artificial IntelligenceValidation & Methodology

Key FindingUsing Hexoskin ECG and acceleration data from 42 hospitalized adults with epilepsy, an offline model detected 56 of 66 focal-to-bilateral tonic-clonic seizures (84.8% sensitivity), with a median of 0.55 false alarms per 24 hours. The study did not test real-time detection at home.

Summary

Forty-two adults with epilepsy wore a Hexoskin connected shirt during simultaneous hospital video-EEG monitoring. An XGBoost pipeline used ECG and accelerometer features to detect 66 focal-to-bilateral tonic-clonic seizures. Respiratory signals were recorded but were not included in the analysis.

HexoskinArticleData Science & Artificial IntelligenceClinical Medicine & SurgeryExercise Physiology & Sports Science

Key FindingThe prototype combining body mass index with Hexoskin measurements assigned 50% of its overall weight to body mass index and 12.5% each to heart rate, breathing rate, minute ventilation and caloric expenditure. The authors reported training with 36 PhysioNet patient records and testing with randomly generated data. No prospective clinical evaluation, diagnostic accuracy or weight-loss benefit was reported, and future self-learning behavior was only proposed.

Summary

The study described a smartphone prototype for individualized activity and overweight-or-obesity risk profiling. It combined Hexoskin activity and vital-sign data with user-reported calorie intake and normalized selected variables using standard limits drawn from best-practice and national guidelines. The resulting weighted score was used to propose daily calorie plans.

HexoskinArticleData Science & Artificial IntelligenceClinical Medicine & SurgeryExercise Physiology & Sports Science

Key FindingThe prototype, which included Hexoskin cardiac, respiratory and activity data, weighted body mass index at 50% of its overall score and heart rate, breathing rate, minute ventilation and caloric expenditure at 12.5% each. Evaluation also used synthetically generated values, which were not participant or wearer records. The paper did not report prospective clinical validation, diagnostic accuracy or weight-loss outcomes; continuous self-learning was described only as future work.

Summary

This development paper described a smartphone prototype that combined body mass index, user-reported information and averaged Hexoskin heart-rate, breathing, ventilation, cadence and calorie data. Three male project students aged 21–23 initially tested one week of garment collection. The app calculated an overall score and selected activity and dietary plans.

HexoskinArticleOccupational Health & Human FactorsNeuroscience, Sleep & CognitionValidation & Methodology

Key FindingMean and maximum heart rate derived from Hexoskin ECG were higher during the stress trial than the control trial, while the heart-rate-variability measure RMSSD showed no significant trial effect. Self-reported distress, challenge and threat also differed across conditions. The preliminary student study had missing physiological records and does not establish performance outside the laboratory or in military populations discussed as a future application.

Summary

Twenty university students completed a laboratory protocol combining cognitive workload, evaluator presence and light cycling. A Hexoskin smart shirt recorded ECG, respiration and movement, while self-reports and interviews measured perceived stress. Technical issues left ECG available for 16 participants, and only cardiovascular measures were compared statistically.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human FactorsValidation & Methodology

Key FindingAcross intervals analyzed using one surgeon’s Hexoskin ECG recordings, surgical-event rates were 47–66% higher in higher heart-rate-variability stress quantiles than in lower quantiles; the strongest association used one-minute RMSSD intervals. Data came from one surgeon, and technical performance was assessed only for gastric bypass and sleeve gastrectomy. The observational association does not establish whether stress caused events or events increased stress.

Summary

One attending surgeon wore a Hexoskin Smart Shirt during 25 consecutive elective laparoscopic and endoscopic procedures. Time-stamped ECG-derived heart-rate variability and procedure video were divided into one-, two- and five-minute intervals. Technical events in the laparoscopic procedures were rated with the Operating Room Black Box and Generic Error Rating Tool.

HexoskinArticleExercise Physiology & Sports ScienceNeuroscience, Sleep & Cognition

Key FindingThe athlete reproduced her individualized breathing pattern between shots with some variability, and heart-rate variability increased during separate paced-breathing exercises. Hexoskin ECG quality during the simulated competition was insufficient for heart-rate-variability analysis. Reported improvements in calm, control and shooting performance came from the athlete and coach; the uncontrolled single case and absent pre-intervention baseline do not establish a training effect.

Summary

This applied case study followed resonant-frequency breathing training in one 45-year-old elite Olympic Trap shooter. During a simulated competition, Hexoskin respiration and acceleration were synchronized to identify shots and examine breathing between them. The athlete reported that the garment did not affect her shooting.

HexoskinArticleNeuroscience, Sleep & CognitionOccupational Health & Human Factors

Key FindingIn a virtual reality study that tracked heart rate and movement with Hexoskin, paired participants reported greater copresence and more positive affect than solo participants. Presence, immersion, flow and state anxiety did not differ significantly, and neither heart rate nor physical movement changed significantly between groups. The convenience sample and field design support an experiential comparison, but they do not establish a physiological effect of shared virtual reality.

Summary

In a field study, 28 adults experienced virtual reality alone or in pairs while wearing a Hexoskin Smart Garment for heart-rate and movement measurement. The experience included a 30-minute active standing phase and an eight-minute passive seated phase. Questionnaires completed after both phases assessed presence, immersion, flow, anxiety, affect and copresence.

HexoskinConference PaperData Science & Artificial IntelligenceNeuroscience, Sleep & CognitionCardiology & ArrhythmiaRespiratory & Infectious Disease

HexoskinArticleExercise Physiology & Sports ScienceNeuroscience, Sleep & CognitionClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingIn the Canadian Bedrest Study, which used Hexoskin for physiological monitoring, 22 of 23 participants completed the bed-rest phase; one withdrew on day 3. Two control participants later withdrew from follow-up after atrial-fibrillation episodes that the ethics committee judged unrelated to the study. This protocol and methodology report coordinates broad physiological, cognitive and biomarker outcomes, but it does not provide comparative results establishing benefit from the exercise countermeasure, and its highly selected healthy cohort limits generalizability.

Summary

The Canadian Bedrest Study used a two-arm randomized, non-blinded design in 23 healthy adults aged 55–65. Its 26-day schedule included 14 days of six-degree head-down bed rest and seven recovery days; half the participants received prescribed resistance and endurance exercise. Hexoskin recorded ECG, respiration and acceleration during designated 24-hour periods.

HexoskinManuscriptNeuroscience, Sleep & Cognition

HexoskinManuscriptNeuroscience, Sleep & Cognition

HexoskinManuscriptNeuroscience, Sleep & Cognition

HexoskinConference AbstractExercise Physiology & Sports ScienceValidation & Methodology

HexoskinArticleRespiratory & Infectious DiseaseData Science & Artificial IntelligenceValidation & Methodology

Key FindingIn the WE SENSE protocol, which includes Astroskin monitoring, the primary planned outcome was within-person change in physiological and activity measures before versus after vaccination. The secondary planned outcome was a binary classifier of within-participant infection state. F1 score, accuracy, sensitivity, and specificity were planned evaluation measures rather than reported performance results.

Summary

The WE SENSE protocol describes a controlled longitudinal trial using wearable sensors around a low-grade influenza-vaccine challenge. The protocol planned to recruit 60 healthy adults, and 58 had enrolled by manuscript review. Astroskin was one of three complementary wearable platforms and the reference device for sleep, electrocardiography, and respiration monitoring. Planned monitoring covered 12 days: 7 days before and 5 days after vaccine administration. Data retrieval and sample analysis were still ongoing at protocol review, so no outcome results were reported.

AstroskinArticleRespiratory & Infectious DiseaseData Science & Artificial IntelligenceValidation & MethodologyExercise Physiology & Sports Science

Key FindingIn the WE SENSE protocol, which includes Astroskin monitoring, the primary planned outcome was within-person change in physiological and activity measures before versus after vaccination. The secondary planned outcome was a binary classifier of within-participant infection state. F1 score, accuracy, sensitivity, and specificity were planned evaluation measures rather than reported performance results.

Summary

The WE SENSE protocol describes a controlled longitudinal trial using wearable sensors around a low-grade influenza-vaccine challenge. The protocol planned to recruit 60 healthy adults, and 58 had enrolled by manuscript review. Astroskin was one of three complementary wearable platforms and the reference device for sleep, electrocardiography, and respiration monitoring. Planned monitoring covered 12 days: 7 days before and 5 days after vaccine administration. Data retrieval and sample analysis were still ongoing at protocol review, so no outcome results were reported.

AstroskinArticleCardiology & ArrhythmiaData Science & Artificial IntelligenceRespiratory & Infectious DiseaseValidation & Methodology

Key FindingAdding night-time and daytime Astroskin/Hexoskin data yielded a reported 24-hour-tolerance ROC-AUC of 0.91, specificity of 0.83 and F1 score of 0.65. The high-sensitivity threshold produced some overprediction at moderate inflammatory-response levels. These results come from a controlled healthy-adult challenge and were not externally validated for clinical diagnosis.

Summary

The WE SENSE trial examined whether machine learning could detect within-person physiological and activity changes surrounding a live attenuated influenza vaccine challenge. A secondary binary classifier predicted an individualized systemic-inflammatory state from wearable and symptom data. Night-time and daytime Astroskin/Hexoskin features were evaluated as part of this controlled challenge. The study does not establish diagnosis of naturally acquired respiratory infection.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingHexoskin cardiorespiratory and energy-expenditure variables did not differ significantly by operative-limb laterality for either right- or left-handed surgeons. During knee arthroplasty, consistently standing on the hand-dominant side was associated with higher HRV and lower minute ventilation, but the authors judged these differences unlikely to be clinically significant. Results from three high-volume surgeons do not establish effects on patient outcomes or broader surgical populations.

Summary

Three high-volume arthroplasty surgeons were monitored during conventional knee and hip arthroplasty in a prospective cohort study. A Hexoskin Smart Vest recorded cardiorespiratory measures and energy expenditure. Operative-limb laterality was not associated with significant differences in the measured variables for right- or left-handed surgeons. Findings about surgeon position came from this small specialist sample and were judged unlikely to be clinically significant.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingIn one surgeon monitored with Hexoskin, robotic-assisted knee arthroplasty took longer but had lower calorie expenditure per minute and lower mean heart rate than conventional surgery in this single-surgeon series. Robotic cases also involved less time in demanding lumbar flexion, while several neck and shoulder measures did not differ. The study did not establish that these physiological and ergonomic differences reduce musculoskeletal pain or injury.

Summary

This study monitored one surgeon during 40 consecutive unilateral knee arthroplasties, comparing 20 robotic-assisted with 20 conventional procedures. Hexoskin Smart Garments worn under surgical scrubs collected cardiorespiratory and caloric-expenditure measures as physiological-stress surrogates alongside wearable ergonomic measurements. Robotic cases took longer but had lower calorie expenditure per minute and lower mean heart rate. Whether these differences reduce surgeon pain or injury remains unknown.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingFor one surgeon wearing Hexoskin, robotic unicompartmental knee arthroplasty involved greater heart rate and total calorie expenditure than conventional instrumentation, while HRV, breathing rate and minute ventilation did not differ significantly. The handheld device reduced neck flexion but increased shoulder abduction. These findings came from one surgeon and do not establish patient outcomes or broader ergonomic benefit.

Summary

The study compared one surgeon during 15 robotic-assisted and 15 conventional unicompartmental knee arthroplasties. Hexoskin garments collected cardiorespiratory and caloric-expenditure data beneath surgical scrubs. Robotic cases took longer and used more calories, while HRV, breathing rate and minute ventilation did not differ significantly. The handheld device reduced neck flexion but increased shoulder abduction, so the expected overall ergonomic and physiological advantages were not realized.

HexoskinArticleNeuroscience, Sleep & CognitionOccupational Health & Human FactorsClinical Medicine & Surgery

Key FindingAcross all students, none of the four mental-health group comparisons reached the paper’s significance threshold. Gender-stratified results differed: several scores were lower for girls in flexible classrooms, whereas boys in flexible classrooms had higher internalizing, inattention/hyperactivity and emotional-symptom scores. Nocturnal heart-rate variability assessed using Hexoskin showed no seating-group difference in the 45-student subsample, and the nonrandomized classroom design limits causal interpretation.

Summary

This study compared 107 Quebec students in three flexible-seating and three fixed-seating classes. Student mental-health measures were analyzed by gender, and teacher data were summarized by group. Girls in flexible classrooms had lower mean scores across the four reported mental-health indicators, while boys had lower school-problem scores but higher scores on the other measures. Nocturnal heart-rate variability in the 45-student subsample, recorded over two consecutive nights with Hexoskin biometric vests, did not differ between seating groups for either gender. The observational classroom comparison does not establish that seating caused the reported differences.

HexoskinArticleExercise Physiology & Sports ScienceRespiratory & Infectious DiseaseValidation & Methodology

Key FindingA custom algorithm for analyzing Hexoskin respiratory-band signals identified more than 99% of flow reversals, with pooled precision and recall of 0.998 and mean lag of 0.018 s (−0.067, 0.104). Breath-cycle time and breathing rate showed low bias and error, while duty cycle showed systematic error and breathing-rate variability had high relative error at low values. Results from 12 runners in a masked laboratory treadmill test do not establish field performance or accuracy for other respiratory measures.

Summary

Twelve runners completed a synchronized incremental treadmill protocol to exhaustion while wearing a Hexoskin smart shirt, RIP sensors and reference spirometry. A custom algorithm filtered and segmented the RIP data to estimate flow reversals and breathing patterns. It identified more than 99% of reference flow reversals, with pooled precision and recall of 0.998 and mean lag of 0.018 s (−0.067, 0.104). Performance was strongest for flow-reversal detection and breathing rate; other breathing-pattern components were less consistent.

HexoskinArticleExercise Physiology & Sports Science

Key FindingIn runners whose respiration and cadence were tracked with Hexoskin, step-adaptive guidance increased indoor and outdoor locomotor-respiratory attachment, although reported outdoor baseline values differ slightly between the Results and table. Outdoor breathing-rate variability and expiratory entrainment stability improved, while the overall inspiratory-stability effect was nonsignificant. The small sequential study lacked randomization, a true no-sound control and an accepted reference for overall coupling accuracy.

Summary

Researchers tested real-time, step-adaptive sound guidance for locomotor-respiratory coupling in 16 novice-to-intermediate female runners. Hexoskin tracked respiration and cadence during indoor and outdoor step-rate-feedback runs followed by individualized breathing-cue trials. Guidance increased step-breath synchronization in the sequential within-subject comparison. The acute, nonrandomized proof of concept had no separate no-sound control and did not assess long-term learning or metabolic benefit.

HexoskinArticleValidation & MethodologyClinical Medicine & SurgeryCardiology & Arrhythmia

Key FindingAgainst 24-hour Holter ECG in 36 children, mean Hexoskin heart-rate accuracy fell from 94.9% in the first 12 hours to 80.0% in the second. A pediatric cardiologist classified 31 of 36 ECG rhythm strips from the shirt; clinically significant slow or fast arrhythmias did not occur, so the study could not assess detection of those rhythms.

Summary

This pediatric cardiology study compared a Hexoskin smart shirt and Corsano CardioWatch with 24-hour Holter ECG. The Hexoskin analysis included 36 children and found mean heart-rate accuracy of 87.4% (SD 11%). Accuracy fell from 94.9% in the first 12 hours to 80.0% in the second 12 hours and declined with greater movement and higher heart rates. A blinded cardiologist classified 31 of 36 Hexoskin rhythm recordings; five were obscured by artifacts or noise. Clinically significant brady- and tachyarrhythmias did not occur, so diagnostic accuracy for those rhythms was not established.

HexoskinPoster / PresentationOccupational Health & Human FactorsNeuroscience, Sleep & CognitionValidation & Methodology

HexoskinConference PaperNeuroscience, Sleep & CognitionData Science & Artificial IntelligenceValidation & Methodology

HexoskinArticleNeuroscience, Sleep & CognitionData Science & Artificial IntelligenceValidation & Methodology

Key FindingThe two-cluster model combining Hexoskin ECG with electrodermal activity and skin temperature from a separate device showed moderate separation with an average silhouette score of 0.40, compared with 0.14 for the three-cluster solution. Under leave-one-participant-out validation, SVM accuracy was 90.9% for binary cluster labels and 73.3% for three-class labels, while the three-class psychophysiological groups were not statistically distinct.

Summary

Forty-four young adults completed a relaxation baseline and four laboratory anxiety-eliciting tasks while Hexoskin recorded ECG and an Empatica E4 recorded electrodermal activity and skin temperature. The analysis used baseline-normalized multimodal features and Gaussian mixture models to form two- or three-level physiological response groups without using self-report as the training label. The two-cluster solution had an average silhouette score of 0.40, while the three-cluster solution scored 0.14. With leave-one-participant-out validation, SVM accuracy was 90.9% for binary cluster labels and 73.3% for three-class labels, but the authors note overlap, modest cluster separation, and difficulty distinguishing intermediate levels.

HexoskinArticleRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceValidation & Methodology

Key FindingAmong the seven participants with paired step-test data, thoracic and abdominal contributions measured with Hexoskin respiratory bands to tidal volume were highly repeatable across low, moderate, and heavy step exercise, with ICCs of 0.87, 0.89, and 0.93. At matched minute ventilation, tidal volume and breathing rate showed no statistically significant difference between ramp and step protocols, but ramp-test compartment signals could not be reliably calibrated to absolute volumes and were analyzed as motion indices.

Summary

Ten healthy adults enrolled in a protocol with one ramp cycling test and two step-test visits, but only seven had paired step-test data; Hexoskin respiratory bands measured thoracic and abdominal motion. A separate turbine-based metabolic system measured tidal volume, breathing rate, minute ventilation, and gas exchange. Relative compartment contributions during repeated step exercise had ICCs of 0.87 at low, 0.89 at moderate, and 0.93 at heavy intensity. At matched minute ventilation, tidal volume and breathing rate showed no statistically significant difference between ramp and step protocols, although calibration limitations restricted ramp-test compartment comparisons to motion indices rather than absolute volumes.

HexoskinConference AbstractCardiology & ArrhythmiaRespiratory & Infectious DiseaseExercise Physiology & Sports Science

HexoskinReportRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceData Science & Artificial IntelligenceValidation & Methodology

HexoskinArticleOccupational Health & Human FactorsData Science & Artificial IntelligenceValidation & Methodology

Key FindingAcross 11 controlled assembly sessions from nine participants, both adaptation strategies reduced mean squared error between the initial theoretical fatigue model and the data-driven estimate. The evaluation is preliminary: Hexoskin access was limited to heart rate and breathing rate, the cohort was small, and repeated sessions within each participant were insufficient for within-person validation.

Summary

Nine project-consortium participants completed 11 controlled manual-assembly sessions after a 20-minute rest, with approximately 1.0–1.3 hours of work per session. Hexoskin supplied 1 Hz heart rate and breathing rate, Xsens supplied 100 Hz motion data for posture and RULA scoring, and Borg ratings captured perceived fatigue every 5–6 minutes. The hybrid framework used a data-driven fatigue estimate to calibrate either the theoretical model’s fatigue rate or its work/rest input signal, reducing session-level mean squared error in both approaches. The small cohort, limited repeated sessions, literature-derived membership functions, and access to only two Hexoskin metrics make the findings preliminary rather than evidence of a deployable production-planning system.

HexoskinConference PaperOccupational Health & Human FactorsData Science & Artificial Intelligence

HexoskinArticleRespiratory & Infectious DiseaseValidation & Methodology

Key FindingChest and abdominal movement measured with Hexoskin was in phase at lower and preferred breathing frequencies, then gradually shifted toward participant-specific thoracic-leading, abdominal-leading, or anti-phase patterns as breathing frequency increased. The study found no abrupt phase transition and did not measure inhaled or exhaled air volume; harmonica breathing, discarded trials, and the limited 7.5–120 bpm range constrain interpretation.

Summary

Sixteen healthy adults completed paced voluntary breathing at 7.5, 15, 30, 60, and 120 breaths per minute while Hexoskin chest and waist strain gauges recorded thoracic and abdominal movement. Relative phase was near in phase at preferred breathing frequency, with a mean angle of −5.58 ± 12.63°, and the absolute phase angle moved farther from in phase as breathing frequency increased. At 120 breaths per minute, seven participants remained in phase, while two showed thoracic-leading, five abdominal-leading, and two anti-phase patterns. The changes were gradual rather than an abrupt phase transition, and the harmonica task, missing trials, unmeasured respiratory volume, and limited frequency range restrict generalization.

HexoskinPoster / PresentationCardiology & ArrhythmiaOccupational Health & Human FactorsClinical Medicine & SurgeryValidation & Methodology

AstroskinReviewCardiology & ArrhythmiaData Science & Artificial IntelligenceValidation & Methodology

Key FindingThe review identifies motion artifacts and scarce labeled examples of abnormal rhythms as obstacles to automated ECG analysis. It does not report accuracy for an Astroskin arrhythmia detector.

Summary

This review examines how ambulatory ECG systems detect abnormal heartbeats and rhythms. It discusses Astroskin as an example of a garment that could combine ECG with movement and breathing signals when investigating motion artifacts.

AstroskinArticleCardiology & ArrhythmiaData Science & Artificial IntelligenceValidation & Methodology

Key FindingOn selected ECG segments from one participant monitored with Astroskin and VIVALNK, the proposed method had lower mean squared error than the comparison algorithms and higher correlation on most examples; processing across more than 600 ten-second segments averaged 1.5201 seconds per segment. The evaluation used recordings from one healthy 32-year-old man, so larger testing in healthy people and cardiac patients is still needed.

Summary

This single-participant methods study evaluated an automatic algorithm for reducing baseline wander and motion artifacts in ambulatory electrocardiogram recordings. Astroskin and VIVALNK devices recorded lead II ECG and accelerometer signals during daily activities; the methods report one hour per device, while the results later state eight hours of acquisition. The pipeline classified usable 10-second segments, combined ECG-derived noise with accelerometer data, and selected examples showed lower mean squared error across the shown examples and higher correlation in most examples than the comparison algorithms. The findings are preliminary because only the author's data were tested, and the method could not fully reconstruct P and T waves when noise contaminated them.

HexoskinConference PaperCardiology & ArrhythmiaRespiratory & Infectious DiseaseExercise Physiology & Sports Science

HexoskinArticleRespiratory & Infectious DiseaseNeuroscience, Sleep & CognitionExercise Physiology & Sports Science

Key FindingIn varsity athletes monitored with Hexoskin, diaphragmatic breathing produced a lower breathing rate and higher tidal volume than paced breathing and produced lower heart rate and higher heart rate variability than progressive muscle relaxation during the intervention. These immediate differences did not carry into the cognitive stress task 10 minutes later, so the study does not establish a lasting stress-response benefit.

Summary

This randomized laboratory study enrolled 76 varsity athletes and compared deep diaphragmatic breathing with progressive muscle relaxation after a common paced-breathing phase. Fitted Hexoskin Biometric Shirts captured ECG, respiration rate, tidal volume, heart rate, and RR intervals across baseline, intervention, and a later cognitive stress task. During the intervention, diaphragmatic breathing produced lower respiration rate, higher tidal volume, lower heart rate, and higher heart rate variability than progressive muscle relaxation; compared with paced breathing, it changed respiration rate and tidal volume but not heart rate or heart rate variability. The group differences did not persist during the cognitive challenge, and interpretation is limited by the absence of an inactive control, the single-session design, baseline heart rate variability imbalance, and unmatched breathing rates.

HexoskinArticleRespiratory & Infectious DiseaseNeuroscience, Sleep & Cognition

Key FindingIn undergraduates monitored with Hexoskin, diaphragmatic breathing increased heart rate variability from baseline, while mindful breathing did not reach statistical significance; the between-group difference was present during the video but absent one week later. A spirituality-by-condition interaction was only marginally significant at p = 0.06, so the proposed moderating role of spirituality remains tentative.

Summary

This randomized study assigned 48 University of Pennsylvania undergraduates to video-guided mindful breathing or diaphragmatic breathing. Hexoskin recordings provided respiration rate, tidal volume, continuous ECG, RR intervals, and SDNN heart rate variability at baseline, during the intervention, and at a one-week follow-up. Both groups reduced respiration rate and increased tidal volume, but only the diaphragmatic-breathing group showed significant heart rate variability increases during the video and at follow-up. Seven participants did not return, technical failures caused missing data, and the small selective sample and brief intervention limit generalizability.

HexoskinAcademic ThesisData Science & Artificial IntelligenceValidation & Methodology

A Middleman Application for Cross-Platform Data Collection from Hexoskin Smart Vest

Key FindingThe Android prototype maintained stable Hexoskin Bluetooth communication and one-to-one agreement among live logs, local storage, and Firebase during the reported short tests. These results support the software pipeline's data-preservation function, but they do not validate the physiological accuracy of the Hexoskin measurements or clinical performance.

Summary

This master's thesis developed the Boracle Middleman mobile application to collect, standardize, and store data from wearable devices. Hexoskin served as the direct Bluetooth Low Energy testbed on Android, streaming heart rate, respiration rate, step count, activity level, and cadence to local logs and a Firebase-backed pipeline. Tests covered 1-, 5-, and 15-minute sessions and separate five-minute sitting, walking, and jogging trials; log, local, and Firebase records matched, with no data loss observed. The work validates prototype data transport and storage under controlled developer testing, without a clinical cohort, physiological-accuracy comparison, completed iOS Hexoskin integration, or long-duration field validation.

HexoskinArticleNeuroscience, Sleep & CognitionClinical Medicine & Surgery

Key FindingPeriocular warming shortened EEG-measured sleep-onset latency by 8.5 minutes relative to sham, below the cited 10-minute clinical-significance threshold. The foot-proximal skin-temperature gradient was associated with objective sleep-onset latency (R²=0.046); Hexoskin-derived normalized high-frequency heart rate variability increased relative to baseline but was not retained in the final regression model.

Summary

This single-blind crossover study analyzed 38 adults with sleep difficulty or insomnia disorder after one of 39 enrolled participants withdrew at baseline. Participants completed a two-day baseline and two seven-night home intervention periods using warming and sham eye masks, with separate portable EEG and iButton skin-temperature measurements and Hexoskin ECG-derived heart rate variability. Warming shortened objective sleep-onset latency relative to sham, while foot-proximal temperature gradient and normalized high-frequency heart rate variability increased relative to baseline. Foot-proximal temperature gradient was associated with objective sleep-onset latency, but the regression explained only 4.6% of its variance and did not retain the heart rate variability measures. The 8.5-minute reduction was below the cited 10-minute clinical-significance threshold, and interpretation is limited by the single-blind home design, portable EEG definition, and Kao Corporation funding.

HexoskinAcademic ThesisCardiology & ArrhythmiaRespiratory & Infectious DiseaseExercise Physiology & Sports ScienceValidation & Methodology

Key FindingDuring maximal treadmill exercise, correlations between Hexoskin and Cosmed ranged from 0.97 to 0.99 for heart rate and 0.89 to 0.99 for respiratory rate, but only 0.16 to 0.37 for ventilation. The results distinguish tracking breathing frequency from estimating the volume of air moved.

Summary

This master’s thesis compared Hexoskin with a Cosmed reference system during treadmill exercise in healthy, physically active adults. The maximal exercise analysis included 28 participants, and 27 contributed to the submaximal analysis. During maximal exercise, heart rate and respiratory rate correlated strongly with the reference, while estimated ventilation correlated weakly.

HexoskinArticleExercise Physiology & Sports ScienceData Science & Artificial IntelligenceValidation & Methodology

HexoskinAcademic ThesisExercise Physiology & Sports ScienceData Science & Artificial IntelligenceValidation & Methodology

Key FindingFor the feature subset labelled “Hexoskin,” the lowest reported RMSE was 1.26 with Random Forest in leave-one-subject-out testing and 1.35 with a multilayer perceptron in leave-one-task-out testing. These results concern a reanalysed sensor configuration and should not be presented as a direct validation of Hexoskin hardware.

Summary

Olschewski reanalysed a dataset from ten healthy adults to compare machine-learning approaches for estimating energy expenditure. One feature configuration, labelled “Hexoskin,” combined waist acceleration, heart rate, breathing frequency, and minute ventilation measured in the underlying multi-device dataset. Random Forest performed best for that configuration when testing on a held-out participant, while a multilayer perceptron performed best when testing on a held-out task. The thesis explores a sensor configuration rather than establishing the accuracy of recordings from a Hexoskin garment, and its limited tuning and preprocessing choices constrain interpretation.

HexoskinConference AbstractCardiology & ArrhythmiaRespiratory & Infectious DiseaseClinical Medicine & Surgery

HexoskinAcademic ThesisClinical Medicine & SurgeryCardiology & Arrhythmia

Étude fondamentale d’un protocole ostéopathique viscéral péricardique sur la physiologie cardiaque

Key FindingThe thesis used repeated wearable ECG measurements to examine physiological responses around manual therapy. Its exploratory subgroup findings require clinical validation and should not be presented as evidence that the protocol treats disease.

Summary

Fatisson investigated changes in heart rate variability around a pericardial osteopathic protocol in healthy adults. The study compared an intervention group with placebo and no-touch groups over four weekly appointments. Hexoskin recorded ECG-derived data for analysis in Kubios-HRV, alongside quality-of-life and mood questionnaires. Exploratory subgroup findings varied with initial LF/HF levels and do not establish a clinical treatment benefit.

HexoskinArticleNeuroscience, Sleep & CognitionData Science & Artificial IntelligenceValidation & Methodology

Key FindingAdding sleep-onset and sleep-offset corrections to a model using Hexoskin recordings increased three-state agreement with polysomnography from 70.2% to 77.4%; sleep/wake agreement reached 90.8%. The authors caution that findings from healthy young sleepers cannot be generalized to pathological conditions.

Summary

Pion-Massicotte and colleagues developed a sleep-classification algorithm using ECG, respiratory movement, and accelerometry recorded with Hexoskin. Young healthy adults were monitored for two consecutive nights in a sleep laboratory alongside polysomnography. The corrected algorithm was evaluated on 35 recordings and achieved 77.4% agreement for wake, NREM, and REM classification. The study supports further investigation in healthy young sleepers; participant counts differ between the abstract and Methods, and the results should not be extrapolated to sleep disorders.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human FactorsValidation & Methodology

Key FindingSeven clinicians evaluated the realism and usability of a virtual-reality airway-training system that included a Hexoskin vest. The paper reports participant feedback and system calibration, but no physiological results from the vest.

Summary

Jain and colleagues developed a virtual-reality airway-training system that combined physical airway models and procedural sensors with a Hexoskin vest. Ten subjects took part in positional calibration, and seven clinicians evaluated realism and usability. The paper reports calibration and survey results but gives no physiological measurements from the vest.

HexoskinReportOccupational Health & Human Factors

HexoskinArticleOccupational Health & Human FactorsRespiratory & Infectious DiseaseValidation & Methodology

Key FindingCombining dust concentration with inhalation rate allowed the authors to model exposure for a worker in a high-exposure zone. Hexoskin contributed respiratory input to the model; it did not measure particulate matter, and the study did not test clinical health outcomes.

Summary

Priyanka and colleagues studied how dust exposure varies across construction activities and materials in a laboratory chamber. Optical particle counters measured PM1, PM2.5, and PM10 at two heights, while Hexoskin supplied inhalation-rate data for a hypothetical worker risk assessment. The results informed proposed locations for dust-control measures based on the activity and the areas of highest exposure. Field testing and assessment before and after installing those controls remain future work.

HexoskinAcademic ThesisOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

Key FindingBlue debriefing light did not produce a statistically significant stress-recovery advantage over fluorescent light in the available results.

Summary

Emergency-department clinicians wore Hexoskin before, during, and after a simulated clinical event in a randomized comparison of blue and fluorescent debriefing light. The available thesis evidence reports a simulation-related stress response but no statistically significant difference between lighting groups in stress level or recovery duration.

HexoskinArticleValidation & MethodologyRespiratory & Infectious Disease

Key FindingHexoskin ranked as the most suitable wearable reviewed, but no available monitor met all six requirements for ambulatory inhalation topography without further development.

Summary

Researchers compared wearable respiratory monitors against six criteria for measuring how people inhale outside the laboratory. Hexoskin ranked highest among the devices reviewed. Even so, none met all six criteria without modification.

HexoskinPreprintValidation & MethodologyRespiratory & Infectious Disease

Key FindingCalibrated Hexoskin respiratory signals supported natural-environment inhalation topography across 4,733 analyzed breathing cycles.

Summary

Ten adult tobacco users enrolled in a three-day study of cigarette or hookah use in daily life. The researchers calibrated Hexoskin respiratory bands against a spirometer each day. Seven participants completed the study, yielding 4,733 breathing cycles for analysis of inhaled volume, flow and timing.

HexoskinArticleValidation & MethodologyRespiratory & Infectious Disease

Key FindingHexoskin showed the strongest linearity and breath-hold stability among the three wearable respiratory monitors tested on the benchtop simulator.

Summary

Researchers tested Hexoskin and two other wearable respiratory monitors on a mechanical chest simulator. Over five repeated trials, the Hexoskin abdominal band produced the most linear response to chest expansion. Its signal also showed virtually no decay during simulated breath holds.

HexoskinConference PaperData Science & Artificial IntelligenceRespiratory & Infectious Disease

AstroskinConference PaperOccupational Health & Human FactorsData Science & Artificial Intelligence

Key FindingThe case study shows how physiological data could inform individualized monitoring. It reports no astronaut cohort and does not test Astroskin performance in space.

Summary

Researchers illustrated a proposed framework for assessing astronaut countermeasures with data from 90 firefighter students in cold-stress training. The students wore Hexoskin shirts; Astroskin Bio-Monitor was discussed as a possible spaceflight application.

AstroskinArticleRespiratory & Infectious DiseaseData Science & Artificial IntelligenceCardiology & Arrhythmia

Key FindingRecovery heart-rate variability from Hexoskin and Astroskin produced the study's strongest immune-response model, with AUC 0.73.

Summary

In a 12-day influenza-challenge study, 55 healthy adults wore Hexoskin or Astroskin garments during repeated three-minute stair tests and two-minute recoveries. A model using heart-rate variability during recovery predicted a blood-biomarker immune-response score with an area under the ROC curve of 0.73. Models based on measurements taken during activity performed less well.

HexoskinAcademic ThesisNeuroscience, Sleep & CognitionValidation & MethodologyData Science & Artificial Intelligence

Key FindingAcross 40 nights in healthy young adults, the Hexoskin-based classifier had mean agreement of 77.54% with polysomnography for wake, non-REM and REM sleep. Mean kappa was 0.54, indicating moderate agreement beyond chance.

Summary

This master’s thesis tested whether a classifier using Hexoskin recordings could distinguish wakefulness, non-REM sleep and REM sleep against polysomnography. Twenty-one healthy young adults participated in a two-night protocol; excluding one participant’s recordings left 40 nights for analysis. The best corrected classifier reached mean agreement of 77.54%, with a mean kappa of 0.54.

AstroskinHexoskinConference PaperOccupational Health & Human FactorsCardiology & ArrhythmiaNeuroscience, Sleep & Cognition

The Astroskin, a Wearable Body Metrics Garment to Measure Physiological Parameters in Non-Clinical Environments: Health and Spaceflight Missions Applications

Key FindingAll four expedition users reported no major discomfort and found the garment easy to use. The field test assessed prolonged wear and usability; it did not establish measurement accuracy or continuous recording for 27 days.

Summary

Robillard and colleagues described Astroskin’s design for monitoring physiology outside the laboratory. During the XPAntarctik expedition, four users wore the garment for a reported total of 27 days. Power constraints limited recording to selected times.

HexoskinConference PaperOccupational Health & Human Factors

HexoskinArticleValidation & MethodologyOccupational Health & Human Factors

Key FindingIn simulations that used Hexoskin ECG to assess residents’ cardiac responses, modified Angoff ratings showed strong panel reliability but did not predict residents' heart rate variability changes during simulation.

Summary

Twenty-five emergency medicine residents wore Hexoskin during clinical simulations. Researchers calculated heart-rate variability from the ECG and compared the residents’ responses with difficulty ratings from five experts. The experts rated scenario difficulty consistently, but their ratings did not predict the residents’ physiological responses.

HexoskinArticleOccupational Health & Human FactorsClinical Medicine & SurgeryCardiology & Arrhythmia

Key FindingIn emergency medicine residents monitored with Hexoskin, different simulation stressors produced different distress responses, but all were associated with similar reductions in heart rate variability from rest.

Summary

Twenty-seven emergency medicine residents wore Hexoskin during six simulated encounters involving medical difficulty, equipment failure or interpersonal conflict. Residents reported different levels of distress depending on the stressor. Heart-rate variability fell from resting levels with all three stressors, but the differences between stressors were not significant.

HexoskinAcademic ThesisValidation & MethodologyData Science & Artificial Intelligence

Key FindingHexoskin respiratory signals supported swallow-event detection, while reliable classification of food structure remained unresolved.

Summary

Ten volunteers wore Hexoskin while swallowing water, custard, apple, bread, and saliva so that thoracic and abdominal respiratory signals could be compared with video-annotated swallow timing. Swallows after food intake were distinguishable from normal breathing, but the study did not identify a universal respiratory pattern that separated all food structures.

HexoskinConference PaperValidation & MethodologyOccupational Health & Human Factors

HexoskinConference PaperNeuroscience, Sleep & CognitionClinical Medicine & SurgeryData Science & Artificial Intelligence

HexoskinArticleOccupational Health & Human FactorsCardiology & Arrhythmia

Key FindingIn caregivers whose ECG and respiration were recorded with Hexoskin, using the transfer-support robot reduced several physiological and perceived indicators of caregiver physical burden compared with wheelchair transfers.

Summary

Eighteen professional caregivers performed transfer tasks using a wheelchair and a transfer-support robot while Hexoskin recorded ECG and respiration. After robot-assisted transfers, caregivers had lower heart rates and reported less physical stress. Heart-rate variability and cardiorespiratory coherence were generally higher with the robot.

HexoskinArticleOccupational Health & Human FactorsClinical Medicine & Surgery

Key FindingTwo arthroplasty surgeons monitored with Hexoskin experienced greater cardiorespiratory strain during total hip arthroplasty than during total knee arthroplasty.

Summary

The methods and results describe two surgeons who wore Hexoskin during 76 consecutive hip and knee replacements. The researchers compared physiological demand across the two operations. Hip replacement was associated with higher surgeon heart rate, minute ventilation and energy expenditure per unit time than knee replacement.

HexoskinArticleOccupational Health & Human FactorsClinical Medicine & Surgery

Key FindingThe two surgeons monitored with Hexoskin showed greater physiological demand during revision joint replacements than during primary joint replacements.

Summary

Two surgeons wore Hexoskin during 70 primary and revision joint replacements. Revision procedures took longer and were associated with higher surgeon heart rate, stress index and energy expenditure. Heart-rate variability was lower during revisions than during primary procedures.

HexoskinArticleValidation & MethodologyData Science & Artificial IntelligenceNeuroscience, Sleep & Cognition

Key FindingThe smartphone’s respiratory-rate estimates had a mean absolute error of 1.6 breaths per minute against Hexoskin. A separate analysis classified three mindfulness skills.

Summary

Researchers used Hexoskin as the respiratory reference for a smartphone system designed to track breathing and progress in mindfulness practice. The smartphone’s respiration estimate differed from the reference by an average absolute error of 1.6 breaths per minute. In a separate analysis, classifiers of mindfulness skill achieved F1 scores of 80% to 84%.

HexoskinAcademic ThesisOccupational Health & Human FactorsCardiology & Arrhythmia

Key FindingHexoskin heart-rate variability measures were incorporated into a structured firefighter resilience assessment during simulated operational training.

Summary

Fifty pre-service firefighters completed baseline and simulated search-and-rescue monitoring with Hexoskin after six additional recruits lost data to garment or battery problems. Women showed higher SDNN and lower mean heart rate than men, but the small female subgroup limits interpretation of the reported sex differences.

HexoskinConference PaperData Science & Artificial IntelligenceExercise Physiology & Sports ScienceValidation & Methodology

HexoskinArticleNeuroscience, Sleep & CognitionExercise Physiology & Sports Science

Key FindingHexoskin recorded significant cardiac and respiratory changes during competitive esports play in 17 analyzed players.

Summary

Nineteen male university esports players enrolled in a study of physiological responses during League of Legends or Overwatch sessions lasting 1.5 to 2.5 hours. Seventeen provided usable Hexoskin recordings. Heart rate, breathing rate and minute ventilation changed significantly from rest; in League of Legends, the winning team had higher values than the losing team.

HexoskinConference PaperOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

HexoskinArticleCardiology & ArrhythmiaData Science & Artificial Intelligence

Key FindingHexoskin recordings provided the physiological reference used to demonstrate respiration-adaptive timing in a low-power neuromorphic pacemaker model.

Summary

Researchers used surface ECG and respiratory recordings from three dogs wearing Hexoskin to develop and test a neuromorphic pacemaker model. The hardware reproduced the target cardiac timing with an average oscillation period of 556 milliseconds. Its pacing also varied with respiratory phase.

HexoskinArticleValidation & MethodologyData Science & Artificial IntelligenceRespiratory & Infectious Disease

Key FindingA public Hexoskin Hx1 dataset served as the benchmark for an ECG-derived respiration algorithm with 3.66 breaths-per-minute mean absolute error.

Summary

Researchers tested an algorithm that estimates respiration from ECG using a public benchmark recorded with Hexoskin Hx1 in 13 people. They tested real-time operation separately in five people using a different sensor. On the Hexoskin benchmark, the mean absolute error was 3.66 breaths per minute and the mean absolute percentage error was 23.69%.

HexoskinAcademic ThesisExercise Physiology & Sports Science

Key FindingBreathing stability was linked to prone shooting success, and inhaling at standing shot release was associated with poorer accuracy.

Summary

Nine junior and 11 professional biathletes wore Hexoskin during roller-ski shooting sessions under increasing physical load. Breathing-pattern stability predicted prone shooting accuracy, while an inhaled pattern at shot release was associated with lower standing accuracy, particularly among junior athletes.

AstroskinPoster / PresentationValidation & MethodologyOccupational Health & Human Factors

Key FindingHeart rate averaged over 15-second blocks correlated strongly with Flexcomp (r = 0.989). Breathing-rate correlation was lower (r = 0.709), and Finapres blood-pressure readings were significantly higher than Astroskin estimates.

Summary

Four adults took part in a rotating-chair motion-sickness experiment on the ground. Researchers compared Astroskin heart and breathing rates with a Flexcomp system and its blood-pressure estimates with Finapres finger measurements.

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingRadiographic knee-girth ratio predicted the surgeon’s energy demand measured with Hexoskin more effectively than patient body mass index.

Summary

One surgeon wore Hexoskin during 185 primary total knee replacements. A higher radiographic knee-girth ratio predicted higher surgeon heart rate, total energy expenditure and energy expenditure per minute. Patient body mass index did not correlate with the surgeon’s energy consumption.

HexoskinReportOccupational Health & Human Factors

HexoskinConference PaperNeuroscience, Sleep & CognitionData Science & Artificial Intelligence

Key FindingAmong four classifiers tested on Hexoskin respiratory recordings from 13 patients, a support vector machine best distinguished seizure from nonseizure periods, averaging 76.7% accuracy.

Summary

Researchers tested whether breathing patterns recorded with a Hexoskin shirt could distinguish seizures from periods between seizures. An epilepsy specialist used video EEG to mark seizure onset and offset without seeing the shirt data. Across 67 seizures in 13 patients, both breathing rate and minute ventilation differed between seizure and nonseizure periods (p < 0.01).

HexoskinArticleNeuroscience, Sleep & CognitionCardiology & Arrhythmia

Key FindingHexoskin monitoring identified an unusual prolonged postictal cardiorespiratory disturbance weeks before sudden unexpected death in epilepsy in this single-patient report.

Summary

One patient who later died from sudden unexpected death in epilepsy and two living controls wore Hexoskin during epilepsy monitoring. Weeks before the death, one seizure was followed by prolonged shallow breathing and a slow heart rate; most other recorded seizures were unremarkable. Separate heart-rate-variability analyses used the video-EEG system.

HexoskinArticleOccupational Health & Human FactorsValidation & Methodology

Key FindingUsing heart rate recorded with Hexoskin in aircraft deicing technicians, heart-rate-reserve and age-adjusted approaches detected more threshold exceedances than working heart rate during open-basket deicing work.

Summary

Twelve aircraft ground-deicing technicians wore Hexoskin during open-basket field work. The shirt updated heart rate every second, and 11 operators provided data suitable for analysis. Heart-rate-reserve and age-adjusted thresholds classified strain as excessive more often than the working-heart-rate method.

AstroskinConference PaperData Science & Artificial IntelligenceCardiology & ArrhythmiaValidation & Methodology

Key FindingWith the Train1 setup, mean absolute error ranged from 6.3 to 11.3 mmHg for systolic pressure and from 5.2 to 9.8 mmHg for diastolic pressure. The lowest errors came from one participant.

Summary

Can wearable signals estimate blood pressure during daily activities? In three participants, researchers trained personalized models using Astroskin ECG, forehead photoplethysmography and acceleration to estimate systolic and diastolic pressure. They evaluated the estimates separately for each participant.

AstroskinArticleData Science & Artificial IntelligenceCardiology & ArrhythmiaValidation & Methodology

Key FindingAcross participants, blood-pressure models using Astroskin ECG, forehead photoplethysmography and acceleration produced mean absolute error that ranged from 6.21 to 7.58 mmHg for systolic pressure and from 5.03 to 6.16 mmHg for diastolic pressure, depending on the model. All models met the reported AAMI bias requirement, while only selected models met the systolic standard-deviation requirement. The personalized results from five young healthy participants do not establish population-level or clinical performance.

Summary

This study developed personalized cuffless blood-pressure models for activities of daily living in five healthy adults. Astroskin supplied ECG, forehead photoplethysmography and acceleration inputs to nonlinear autoregressive neural-network models and comparator pulse-arrival-time models. Estimates were compared with Portapres finger blood pressure during six and a half hours of data collection per participant. Two participants returned about four months later, and one was tested again at six months.

AstroskinArticleData Science & Artificial IntelligenceValidation & MethodologyCardiology & Arrhythmia

Key FindingCombining blood-pressure estimates from models using Astroskin signals improved blood-pressure and prediction-interval precision by up to 24%; the result was obtained in five healthy participants after participant-specific training.

Summary

This study evaluated uncertainty-aware fusion of three cuffless blood-pressure models in five healthy participants. Astroskin supplied ECG, PPG, and pulse-arrival-time inputs, while Portapres finger blood pressure served as the reference. Each participant completed 6.5 hours of monitoring. Fusion improved blood-pressure and prediction-interval precision by up to 24%, but the small healthy sample and participant-specific training limit generalizability.

HexoskinAcademic ThesisOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

Key FindingA brief sleep and stress intervention was associated with longer sleep and improved shift heart-rate variability among emergency clinicians.

Summary

Fifty-nine physicians and nurses in two emergency departments participated in a cluster-randomized sleep and stress management study that used Hexoskin ECG during work shifts. The thesis reports a 9% improvement in work-hour heart-rate variability and about 15 additional minutes of sleep per night in the intervention group, with stronger cardiac effects among caregivers with moderate burnout.

HexoskinConference PaperOccupational Health & Human FactorsExercise Physiology & Sports Science

HexoskinAcademic ThesisOccupational Health & Human FactorsExercise Physiology & Sports Science

Key FindingTask-synchronized Hexoskin monitoring identified specific open-basket deicing activities associated with elevated cardiac workload.

Summary

Twelve aircraft-deicing technicians wore Hexoskin during 15 hours and 52 minutes of open-basket field recording, with heart rate synchronized to video-coded work tasks. Heart rate exceeded the individual steady-state threshold for a cumulative 39 minutes, and fluid spraying, vehicle and basket movement, and control tasks were among the most demanding activities.

AstroskinArticleValidation & MethodologyNeuroscience, Sleep & Cognition

Key FindingAstroskin showed strong agreement for several ambulatory physiological measures, while blood-pressure precision and sleep-staging availability remained important limitations.

Summary

Researchers compared Astroskin with reference methods for oxygen saturation, temperature, heart rate, systolic blood pressure, respiration and sleep at home. Several physiological measures showed strong agreement, but systolic blood pressure had low precision. The system detected sleep successfully, yet produced sleep stages in only 41% of attempts.

HexoskinConference AbstractValidation & Methodology

Feasibility of the Hexoskin™ Biometric Shirt Use in a Focused Breathing Study

HexoskinConference AbstractRespiratory & Infectious DiseaseValidation & Methodology

HexoskinArticleValidation & MethodologyData Science & Artificial Intelligence

Key FindingHexoskin reference measurements supported validation of contact-free respiration sensing at distances up to six metres.

Summary

Fifteen participants wore Hexoskin as the respiratory reference for LASense, a contact-free acoustic system. At four metres from the smart speaker, LASense had a median error of 0.21 breaths per minute. Errors remained low across several room and noise conditions.

HexoskinArticleNeuroscience, Sleep & CognitionData Science & Artificial Intelligence

Key FindingTransitions among individualized Hexoskin-derived physiological states were associated with near-term changes in negative affect and impulsivity.

Summary

Ninety-three adults wore Hexoskin for seven days and reported their feelings and behavior throughout the day. Individual models identified six to 13 physiological states per person. Where the states matched interpretable patterns, selected transitions predicted higher negative affect or impulsivity.

HexoskinConference AbstractCardiology & ArrhythmiaRespiratory & Infectious Disease

HexoskinAcademic ThesisValidation & MethodologyNeuroscience, Sleep & CognitionData Science & Artificial Intelligence

Key FindingHexoskin reference data supported validation of wrist-based inter-beat interval and breathing estimates during adaptive biofeedback training.

Summary

Ten participants completed respiratory-sinus-arrhythmia biofeedback training while a wrist-worn BreathCoach system was evaluated against Hexoskin ECG and respiration. BreathCoach averaged 9.6 milliseconds of inter-beat-interval error and 0.61 seconds of breathing-cycle-duration error, and the adaptive training improved several biofeedback and recovery measures.

HexoskinAcademic ThesisRespiratory & Infectious DiseaseValidation & Methodology

Key FindingHexoskin tracked respiratory volume changes well, while reliable dynamic-hyperinflation classification remained limited.

Summary

A two-phase study compared Hexoskin respiratory inductance signals with conventional lung-function equipment in 31 healthy adults and 11 patients, 10 of whom had COPD. Signal correlations were generally strong, but agreement for detecting dynamic hyperinflation was low, with Cohen kappa values of about 0.2 to 0.3.

HexoskinArticleOccupational Health & Human FactorsExercise Physiology & Sports Science

Key FindingGreater harvesting automation was associated with lower operator heart rate measured with Hexoskin than fully manual harvesting for both monitored operators.

Summary

Two wild-blueberry harvester operators wore Hexoskin during manual, semi-automated and fully automated harvesting. Compared with fully manual harvesting, full automation reduced average heart rate by 13.83% in the experienced operator. The corresponding reduction in the newer operator was 19.03%.

HexoskinArticleNeuroscience, Sleep & CognitionClinical Medicine & Surgery

Key FindingIn a household-task study using Hexoskin for physiological monitoring, simulated macular degeneration affected task performance and psychological state without a significant change in the monitored physiological measures.

Summary

Twenty-four normally sighted adults performed household tasks with normal vision and with simulated age-related macular degeneration. Researchers used Hexoskin to track physiology, time and steps, and also assessed task performance and distress. Simulated impairment slowed task completion, increased anxiety and distress, and reduced engagement, without significant changes in heart rate, heart-rate variability or respiratory rate.

HexoskinConference PaperData Science & Artificial IntelligenceNeuroscience, Sleep & CognitionOccupational Health & Human Factors

AstroskinArticleOccupational Health & Human FactorsValidation & Methodology

Key FindingThe researchers estimated that fitting Astroskin took less than 10 minutes per participant. Participants reported no discomfort while wearing the shirts. Full quantitative validation of Astroskin data quality was deferred to a separate publication.

Summary

This short report described wearable physiological monitoring in pilot healthcare-simulation studies with paramedicine students. The team added Astroskin monitoring to its established use of Hexoskin in simulation research. Both shirts streamed to an iPad and uploaded data for dashboard analytics before further analysis. The pilot evaluated deployment and participant comfort rather than data quality. The report did not state a participant count.

HexoskinArticleOccupational Health & Human FactorsNeuroscience, Sleep & CognitionExercise Physiology & Sports Science

Key FindingIn a crossover simulation study using Hexoskin to monitor paramedic trainees and recent graduates, occupational activity increased physiological demand and perceived exertion before simulation without a clear overall reduction in neurocognitive test performance.

Summary

Eleven final-year paramedicine students or recent graduates wore Hexoskin during simulations after an occupational carrying task and after seated rest. Heart rate, respiratory rate, movement and perceived exertion were higher after the carrying task. The two neurocognitive tests showed no overall intervention effect.

HexoskinAcademic ThesisOccupational Health & Human FactorsData Science & Artificial Intelligence

Key FindingIn a repetitive lifting experiment, Hexoskin measurements and task characteristics were used to model self-rated exertion separately for male and female participants.

Summary

This thesis investigated whether wearable measurements could help estimate perceived exertion during repetitive lifting in a simulated manufacturing setting. The analysis paired Hexoskin measurements with participants’ Borg exertion ratings and considered task duration, interval and distance alongside breathing rate, heart rate, minute ventilation and activity. The researcher fitted separate regression models for male and female participants and compared model fit on training and test data.

HexoskinConference AbstractValidation & MethodologyRespiratory & Infectious Disease

HexoskinPaperValidation & MethodologyRespiratory & Infectious Disease

HexoskinArticleOccupational Health & Human FactorsExercise Physiology & Sports Science

Key FindingIn eight participants exercising in protective clothing with Hexoskin monitoring, upper-body and whole-body liquid cooling both reduced heat-related cardiovascular strain and improved exercise completion in protective clothing.

Summary

Eight participants exercised in protective clothing with no cooling, upper-body cooling or whole-body cooling while Hexoskin monitored heart rate. Both cooling conditions reduced cardiovascular strain, and every participant completed all three exercise bouts with cooling. Without cooling, no participant began the third bout.

HexoskinArticleOccupational Health & Human FactorsCardiology & ArrhythmiaNeuroscience, Sleep & Cognition

Key FindingIn simulated rescue work monitored with Hexoskin ECG, the breathing-apparatus conditions increased completion time and perceived exertion and delayed parasympathetic recovery relative to protective clothing alone. Breathing frequency was lower with the full apparatus than in the other conditions, while mean and peak heart rate remained high across interventions. Six participants had a recording failure in one condition, and the simulated protocol does not establish outcomes in real fires.

Summary

This randomized crossover study compared firefighter workload and recovery under protective clothing alone, self-contained breathing apparatus with the cylinder, and the full breathing apparatus. Thirty-four firefighters participated, and complete physiological data were analyzed for 28. Hexoskin continuously recorded ECG, chest motion and acceleration during a fitness test and three simulated rescue interventions. Heart rate, breathing frequency, movement, heart-rate recovery and heart-rate variability were analyzed.

HexoskinArticleOccupational Health & Human FactorsNeuroscience, Sleep & CognitionCardiology & Arrhythmia

Key FindingHeart rate measured with Hexoskin was higher and several parasympathetic heart-rate-variability measures were lower during both on-call conditions than during the control night. Total sleep time was shortest on the night with the simulated intervention. The small workplace study relied on wearable-derived sleep measures and one simulated intervention rather than polysomnography or real-fire outcomes.

Summary

This crossover field study compared cardiac-autonomic activity and sleep across three nights in firefighters: a home control night, an on-call night without an intervention and an on-call night with a simulated intervention. Thirteen firefighters participated, and 12 were included in the cardiac and sleep analyses. Hexoskin recorded ECG, respiration and movement and supplied nocturnal heart-rate-variability and dashboard sleep measures. The simulated intervention occurred about 96 minutes after bedtime and lasted 4.2 minutes on average.

HexoskinArticleOccupational Health & Human FactorsCardiology & ArrhythmiaNeuroscience, Sleep & Cognition

Key FindingIn simulated rescue interventions monitored with Hexoskin ECG, nighttime sound alarms produced greater heart-rate reactivity than daytime sound alarms, while the vibrating alarm reduced reactivity relative to the nighttime sound alarm. Recovery after the nighttime sound-alarm condition was slower than after the daytime condition on several measures. The findings come from 16 male firefighters in simulated interventions and do not establish outcomes during real emergencies.

Summary

This workplace simulation study compared day and night rescue interventions in 16 male firefighters. Hexoskin recorded ECG during an intermittent fitness test and three rescue conditions: daytime sound alarm, nighttime sound alarm and nighttime vibrating alarm. The analysis covered alarm mobilization, intervention workload, heart-rate recovery and post-exercise heart-rate variability. Sessions were separated by three to 14 days.

HexoskinArticleValidation & MethodologyOccupational Health & Human FactorsCardiology & Arrhythmia

Key FindingIn analyses of recovery from Hexoskin ECG recordings, the sixth one-minute segment showed good-to-excellent agreement with the five-minute criterion across the tested conditions, whereas earlier segments were less consistent. Higher-fitness firefighters completed rescue tasks faster, but post-exercise LnRMSSD did not differ between fitness groups. Results from 24 simulated-rescue participants require confirmation in larger operational cohorts.

Summary

This study examined post-exercise parasympathetic recovery and physical fitness in 24 firefighters. Hexoskin recorded ECG during an intermittent fitness test, three simulated rescue interventions and pre- and post-exercise rest. One-minute post-exercise RMSSD segments were compared with a five-minute criterion window, and firefighters were divided at the median fitness-test velocity. The shirt was a monitoring tool rather than the device under evaluation.

HexoskinArticleOccupational Health & Human FactorsExercise Physiology & Sports ScienceValidation & Methodology

Key FindingPerceived exertion was the only workload measure that distinguished all three equipment conditions; post-exercise Hexoskin ECG-derived heart-rate variability did not differ, and the results came from simulated rescues in 22 male firefighters.

Summary

This randomized within-subject study compared firefighter workload under three protective-equipment conditions in 22 male firefighters. Hexoskin supplied ECG and hip-acceleration data during simulated rescue interventions. Heart-rate-based and perceived-exertion workloads were higher with the cylinder and full breathing apparatus than with protective clothing alone, while acceleration-based workload was lower and post-exercise heart-rate variability did not differ. The simulated all-male study does not establish operational outcomes in real fires.

HexoskinArticleOccupational Health & Human FactorsClinical Medicine & SurgeryValidation & Methodology

Key FindingThe multimodal evaluation identified ventilator-specific differences in clinician stress and usability, with Hexoskin contributing cardiorespiratory monitoring.

Summary

Twenty intensive-care physicians tested multiple ventilators while Hexoskin tracked heart rate, respiratory rate and changes in thoracic volume. Physiological activation, usability and task performance varied between ventilators. The paper did not report the shirt’s heart-rate results separately.

HexoskinReviewOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

AstroskinConference AbstractOccupational Health & Human FactorsExercise Physiology & Sports ScienceValidation & Methodology

Key FindingCombining ECG leads increased the usable recording time from 92% to 97% in flight and from 96% to 98% before flight. High-force movement fell from just over 2.5 hours per day before flight to about 50 minutes per day in orbit.

Summary

Astronauts wore the Canadian Space Agency’s Bio-Monitor shirt before launch and aboard the International Space Station. Researchers analyzed ECG quality and used accelerometry to compare time spent in high-force movement on Earth and in orbit.

HexoskinArticleOccupational Health & Human FactorsClinical Medicine & Surgery

Key FindingIn multilevel models, the operating surgeon’s Hexoskin-estimated energy expenditure was higher for mild valgus (P = 0.026) and moderate/severe valgus (P = 0.002) versus mild varus. Heart rate (P = 0.16) and minute ventilation (P = 0.23) did not differ across the four groups.

Summary

The study compared surgeon physiologic demand across coronal plane deformity groups in primary total knee arthroplasty. The study included eighty-five patients undergoing primary, elective cemented TKA, stratified by a 10 degree hip-knee-ankle angle threshold into mild varus, moderate/severe varus, mild valgus, and moderate/severe valgus groups. A Hexoskin smart garment worn by the operating surgeon recorded heart rate, minute ventilation, and energy expenditure during primary TKA cases. Energy expenditure differed across the four groups (P < 0.001). The garment itself was not evaluated.

HexoskinStudy ProtocolNeuroscience, Sleep & CognitionData Science & Artificial Intelligence

HexoskinReportValidation & Methodology

HexoskinConference PaperData Science & Artificial IntelligenceValidation & Methodology

HexoskinArticleNeuroscience, Sleep & Cognition

Key FindingHigher Hexoskin minute ventilation (B = 0.036, SE = 0.015, p = .014) and breathing rate (B = 0.052, SE = 0.022, p = .018) predicted greater high-arousal positive affect 30 minutes later.

Summary

The study investigated within-person time-lagged associations between respiratory parameters and subsequent high-arousal positive affect. Participants included adults aged 18-30 years with remitted major depressive disorder or no lifetime psychiatric history. The Hexoskin smart shirt was worn during a 7-day EMA period to continuously record respiration via RIP sensors. Breathing rate and minute ventilation predicted greater high-arousal positive affect 30 minutes later. The shirt was not evaluated against a reference.

HexoskinAcademic ThesisData Science & Artificial IntelligenceClinical Medicine & Surgery

Early Detection and Self-Management of Long-Term Conditions using Wearable Technologies: An Application of Pre-diabetes Detection and Self-management of Diabetes Type – II

Key FindingThe proposed system reported 91.49% accuracy, 94% sensitivity, and 90% specificity. With only two participants and limited wearable use, those figures should be treated as prototype results rather than evidence of clinical diagnostic performance.

Summary

This thesis proposed a wearable-data system for early prediabetes detection and type II diabetes self-management. Two participants supplied demographic, vital-sign, activity, blood-glucose, and HbA1c data, and one used a Hexoskin vest for real-time monitoring. An adaptive neuro-fuzzy inference model and individualized interpretation engine combined the available inputs. The thesis reports system-level performance estimates but does not provide a clinical validation cohort.

HexoskinArticleOccupational Health & Human FactorsRespiratory & Infectious Disease

Key FindingDuring resting mask trials that included Hexoskin monitoring, surgical and N95 masks caused small resting metabolic and respiratory changes that remained within expected physiological ranges.

Summary

Forty-two adults completed randomized 15-minute resting trials wearing no mask, a surgical mask or an N95 respirator. Hexoskin recorded heart rate, respiratory rate and minute ventilation. Masks produced small increases in respiratory exchange ratio, minute ventilation and perceived exertion, with the highest perceived exertion in the N95 condition.

HexoskinConference PaperValidation & MethodologyCardiology & Arrhythmia

HexoskinArticleData Science & Artificial IntelligenceValidation & Methodology

Key FindingThe biometric classifier evaluated on Hexoskin ProShirt ECG reported accuracy above 98%, a 0.93% false-acceptance rate, a 3.64% false-rejection rate and a true-positive rate above 96% on the in-house dataset. Results from 11 participants and sample-level model development do not establish performance in larger populations or independent deployments.

Summary

This study evaluated ECG biometric authentication from Hexoskin ProShirt recordings under different physiological conditions. Eleven participants were recorded while standing, sitting, walking and completing uncontrolled activity. After filtering and QRS-based feature extraction, 854 samples were used for training and validation and 300 for testing a quadratic support-vector machine. The study assessed the classifier rather than validating Hexoskin against a clinical ECG reference.

HexoskinArticleData Science & Artificial IntelligenceValidation & Methodology

Key FindingAcross ECG recordings from participants wearing either Hexoskin or HeartIn Fit shirts, a tuned quadratic support-vector machine achieved 97.4% accuracy across the study data; the result comes from 22 healthy participants and short recordings from two different shirt brands.

Summary

This study evaluated ECG biometrics from two commercially available smart-textile shirts under a real-life recording protocol. Twenty-two healthy participants wore either a Hexoskin ProShirt or a HeartIn Fit shirt for at least 15 minutes. Hexoskin recorded ECG at 256 Hz through three textile electrodes. A tuned quadratic support-vector machine achieved 97.4% accuracy, but the small healthy sample and short brand-specific recordings do not establish broader biometric performance.

HexoskinArticleData Science & Artificial IntelligenceValidation & Methodology

Key FindingThe classifier achieved accuracy above 99.63%, a 0.14% false-acceptance rate, a 2.86% false-rejection rate, and a 97.14% true-positive rate. The classifier used approximately 854 datasets for training and validation and 300 datasets for testing. Because the dataset combined Hexoskin and HeartIn shirts, the reported classifier performance was not Hexoskin-specific.

Summary

The study assessed electrocardiogram signals from Hexoskin ProShirt and HeartIn Fit shirts for biometric verification in real-world settings. The study collected data from 22 participants over more than 30 days. Participants wore commercial Hexoskin ProShirt or HeartIn Fit smart-textile shirts. The method combined time-domain preprocessing, QRS feature extraction, and a quadratic support-vector-machine classifier. The sample comprised 22 healthy participants, including 18 men and 4 women.

HexoskinAcademic ThesisValidation & Methodology

Remote vital signs monitoring using a mm-wave FMCW radar

Key FindingRadar estimates closely tracked the Hexoskin reference in the reported single-participant experiments, including a 40-minute back-sleeping trial. The one-person design and controlled geometry prevent generalization to clinical monitoring or diverse sleepers.

Summary

This thesis developed a millimeter-wave frequency-modulated continuous-wave radar method for non-contact monitoring of breathing and heart rate. One healthy male participant was recorded while sitting or sleeping in different positions, with a Hexoskin vest supplying the reference breathing and cardiac signals. Signal processing included phase correction, harmonic analysis, and adaptive filtering for waveform reconstruction. Reported worst-case accuracy was 95% for breathing rate and 94% for heart rate within the study's stated tolerances.

HexoskinConference AbstractValidation & MethodologyExercise Physiology & Sports Science

Validité de critère du textile intelligent Hexoskin pour évaluer la fonction cardiorespiraoire chez une population en santé.

HexoskinArticleNeuroscience, Sleep & CognitionOccupational Health & Human Factors

Key FindingNo significant effect of the virtual-reality training on physiological responses monitored with Hexoskin was detected; with 15 analyzed students, the study had limited power to detect small effects.

Summary

This study tested a mixed-reality virtual-reality training program for music performance anxiety. The study included 15 students (19 enrolled, 4 excluded). The analysis used linear mixed-effects models per metric with group × session interaction, instrument type and gender as fixed effects and subject as random effect. Hexoskin smart shirts were worn during the resting state, baseline and exam sessions as the physiological monitoring tool, recording single-channel ECG and thoracic and abdominal respiration from which heart rate, respiratory rate and variability measures were derived. No significant training effect was detected, but the small sample and sensitivity analysis indicated limited power for small physiological effects.

HexoskinArticleValidation & MethodologyNeuroscience, Sleep & Cognition

Key FindingThe Hexoskin algorithm achieved 65% prone, 75% supine, 94% right lateral decubitus, 81% left lateral decubitus, and 65% sitting/standing classification accuracy compared to VEEG. Excluding intermediate positions (336 h) resulted in 80% prone, 76% supine, and 84% left classification accuracy. Results are based on a sample of 10 patients.

Summary

The study evaluated the Hexoskin shirt for identifying sleep body positions in patients with epilepsy. The study included 10 adult patients admitted to the CHUM epilepsy monitoring unit for at least 48 hours. The Hexoskin built-in accelerometer sleep position classification algorithm was compared second by second with blinded video-EEG annotations. The analysis included 347 hours of sleep from 40 nights.

HexoskinAcademic ThesisData Science & Artificial IntelligenceNeuroscience, Sleep & Cognition

AMD: Analysis of Mood Dysregulation - A Machine Learning Approach

Key FindingThe all-subject random-forest model reported 95.84% accuracy and a ROC area of 99.2%, while the J48 decision tree reported 93.46% accuracy. These results came from a small filtered sample with resampled classes and require external validation before any clinical interpretation.

Summary

This thesis developed a machine-learning pipeline for analyzing mood dysregulation from wearable physiology and self-reports. Forty women receiving treatment for emotional-distress disorders enrolled, while the prediction module ultimately analyzed seven participants after data availability and quality filters. Hexoskin collected ECG, respiration, and activity during approximately two weeks of daily life, and self-reported mood dysregulation supplied the target labels. Decision-tree, random-forest, Bayesian-network, and naive-Bayes models were evaluated after feature engineering and class balancing.

HexoskinConference PaperNeuroscience, Sleep & Cognition

HexoskinConference AbstractOccupational Health & Human FactorsNeuroscience, Sleep & CognitionClinical Medicine & Surgery

Tracking Mental Workload by Multimodal Measurements During Minimally Invasive Surgery Training

HexoskinConference PaperOccupational Health & Human FactorsNeuroscience, Sleep & CognitionClinical Medicine & Surgery

HexoskinStudy ProtocolClinical Medicine & SurgeryValidation & Methodology

HexoskinConference PaperValidation & MethodologyCardiology & ArrhythmiaData Science & Artificial Intelligence

HexoskinAcademic ThesisNeuroscience, Sleep & CognitionData Science & Artificial Intelligence

Détection automatique des crises d'épilepsie par un chandail connecté

Key FindingThe shirt-based system detected 56 of 66 seizures, with median time in false warning of 10 seconds per alarm. Performance was measured in a monitored epilepsy-unit cohort and does not establish effectiveness for unsupervised home use.

Summary

This thesis developed a machine-learning system to detect focal to bilateral tonic-clonic seizures from Hexoskin electrocardiography and accelerometry recorded alongside video-EEG. Of 148 adults recruited in an epilepsy monitoring unit, 42 participants with epilepsy contributed 66 seizures and 8,067.9 hours of recordings to the analysis. An XGBoost classifier combined heart-rate-variability and movement features, with video-EEG serving as the reference. The reported system reached 84.8% sensitivity, a median 0.55 false alarms per 24 hours, and a ROC-AUC of 0.90.

HexoskinAcademic ThesisRespiratory & Infectious DiseaseData Science & Artificial IntelligenceExercise Physiology & Sports Science

Predicting Viral Respiratory Tract Infections Using Wearable Garment Biosensors

Key FindingThe best reported test result combined XGBoost with recovery-period heart-rate-variability features, yielding 0.70 sensitivity, 0.77 specificity, and 0.75 precision. The test set contained only five participants, so the result is preliminary and does not establish clinical infection detection.

Summary

This thesis examined whether physiological signals from Hexoskin and Astroskin garments could help predict vaccine-induced inflammation as a model of viral respiratory infection. Fifty-five healthy adults wore a garment during a 12-day protocol that included repeated three-minute stair-stepping tests before and after live attenuated influenza vaccination. Heart rate, breathing rate, minute ventilation, and RR-interval features were used in XGBoost and random-forest models. In a five-participant test cohort, the best model used heart-rate-variability features from the first recovery minute and achieved an AUC of 0.73.

HexoskinConference PaperOccupational Health & Human Factors

L'origine de l'objectif est-elle importante? Effets motivationnels d'objectifs autodéfinis en production

HexoskinArticleOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

Key FindingAI automation level had a significant effect on error-detection performance: F(2, 27) = 5.65, p = .009. Physical-effort variability was lower with partial than full automation: t = -5.18, p < .001. Neither the LF/HF ratio nor respiration rate differed across automation levels in the Hexoskin monitoring data.

Summary

The study compared no, full, and partial AI decision-selection automation during manufacturing training. A total of 102 participants completed the study, and 100 were retained for analysis. The Hexoskin smart vest recorded electrocardiography, respiration, and acceleration as engagement measures. Outcomes included task performance, error detection, motivation, and cognitive, emotional, and behavioral engagement. The Hexoskin vest served as a monitoring tool and was not evaluated as a device.

HexoskinArticleData Science & Artificial IntelligenceOccupational Health & Human Factors

Key FindingIn simulated driving, the multimodal model incorporating Hexoskin measurements achieved 79% average accuracy and a 75% average F-score in 15 healthy male volunteers; performance outside the two controlled driving scenarios remains untested.

Summary

This controlled study evaluated textile electrodes and a cross-attention model for driver-stress classification. Fifteen healthy male volunteers wore Hexoskin shirts during baseline and stress-driving scenarios lasting about one hour each. The shirts recorded one-lead ECG at 256 Hz and respiration at 128 Hz, while the study also collected vehicle kinematics. The model achieved 79% average accuracy and a 75% average F-score, but the small all-male sample and two controlled scenarios limit generalizability.

HexoskinConference PaperData Science & Artificial IntelligenceOccupational Health & Human FactorsValidation & Methodology

Key FindingA model using Hexoskin ECG to classify distraction during driving achieved weighted F-scores of 0.53 to 0.75 when each driver was tested separately from the drivers used for training.

Summary

Ten licensed drivers wore a Hexoskin shirt that recorded single-lead ECG during baseline driving, texting and calling in a controlled setting. Researchers divided the ECG into 10-, 30- and 60-second segments to train a convolutional neural network to distinguish the three conditions. On the validation set, the model achieved 67.12% average accuracy and a weighted F-score of 0.65.

HexoskinConference PaperData Science & Artificial IntelligenceOccupational Health & Human Factors

Key FindingCombining Hexoskin ECG and respiration early in the model yielded 65.55% accuracy and a 62.77% F1 score for distinguishing normal driving from driving during a phone call.

Summary

Ten male volunteers wore Hexoskin shirts while driving normally and while making phone calls under controlled conditions. The shirts recorded single-lead ECG at 256 Hz and respiration at 128 Hz; the researchers combined the signals in a neural network. They tested the model on one driver at a time, training it on the remaining drivers.

HexoskinArticleRespiratory & Infectious DiseaseExercise Physiology & Sports Science

Key FindingAverage Hexoskin respiratory rate remained stable across 15 weeks in the two consistently monitored participants, with no difference between choir and dance sessions. Spirometry was stable in two participants and improved in one, while quality-of-life improvements were reported for two participants. The very small, incomplete outcome samples make these feasibility findings preliminary.

Summary

This mixed-methods study examined a 15-week singing and dancing program for adults with neuromuscular conditions. Six participants with neuromuscular conditions entered the research study, but each outcome had a smaller and different sample. A modified Hexoskin shirt provided respiratory monitoring during weekly 90-minute sessions and was calibrated to spirometry. Four participants donned the shirt and two wore it consistently across the program.

HexoskinArticleRespiratory & Infectious DiseaseExercise Physiology & Sports Science

Key FindingAcross both groups, participants with BP angle ≤45° reported greater peak dyspnoea than those above 45° (visual analogue scale 86.5 vs 65.5 mm, p = 0.036; Borg 8 vs 4, p = 0.001); the result did not use Hexoskin data.

Summary

This cross-sectional study compared exercise breathing patterns and dyspnoea in adults with well-controlled asthma and healthy adults. The study enrolled 32 adults, 16 with asthma and 16 healthy controls; 30 completed the protocol and were analyzed, 15 per group. The primary outcome was breathing pattern quantified by the new BP angle, arctan of delta VT (cL) over delta BF times FVC between 40 % and 60 % of max load, with a 45 deg cut-off. All subjects wore a Hexoskin vest to measure thoracic and abdominal breathing with respiratory inductance plethysmography, but the authors state these measurements fall outside the scope of this paper. Mean BP angle was similar in asthma and healthy groups (66.8° ± 25.4° vs 72.9° ± 26.4°); these measurements came from the metabolic cart, not Hexoskin.

HexoskinAcademic ThesisOccupational Health & Human FactorsExercise Physiology & Sports ScienceNeuroscience, Sleep & Cognition

Key FindingOn-call nights with an intervention had a mean Hexoskin total sleep time of 281.5 minutes, compared with 432.2 minutes on control nights and 399.5 minutes on on-call nights without an intervention. These small, related studies used simulated or station-based conditions and do not establish causal operational outcomes.

Summary

This thesis examined psychophysiological responses of firefighters during simulated fire-suppression interventions across five related studies. Hexoskin continuously recorded physiological and activity measures during fitness testing, simulated interventions, recovery periods, and selected on-call nights. The studies addressed workload, cardiac parasympathetic recovery, day-versus-night stress, post-exercise heart-rate-variability methods, and sleep during on-call duty. Participant counts differed by study and ranged from 13 to 34 firefighters.

HexoskinArticleValidation & MethodologyExercise Physiology & Sports Science

Reliability and Validity of the Hexoskin Wearable Body Metrics Telemetry Shirt

Key FindingHexoskin calorie estimates showed strong reliability and criterion validity during vigorous exercise but weaker, nonsignificant results during moderate exercise. Breathing-rate validity was high at both intensities, while test-retest reliability was inconsistent and improved after two outliers were removed. The six-person sample and outlier sensitivity make the estimates preliminary.

Summary

This small laboratory study examined the reliability and validity of Hexoskin during moderate and vigorous treadmill exercise. Six healthy trained adults completed 15-minute bouts on nine testing days over three weeks. Calories per minute, breathing rate and heart rate were compared across repeated tests and, where applicable, with a ParvoMedic metabolic cart. The shirt was the device under evaluation.

HexoskinConference AbstractExercise Physiology & Sports ScienceNeuroscience, Sleep & Cognition

Effect of Heat on Psycho-Physiological Responses During a Self-Paced Cycling Time Trial

HexoskinConference PaperData Science & Artificial IntelligenceExercise Physiology & Sports Science

HexoskinArticleExercise Physiology & Sports ScienceValidation & Methodology

Key FindingFor thresholds derived from Hexoskin ECG, the control group repeated-measures correlation was rrm = 0.66 (0.32; 0.84) for HRV0.75 vs VT1 and rrm = 0.82 (0.61; 0.92) for HRV0.5 vs VT2 (both P < 0.001). The analysis was based on 68 of 88 planned test records.

Summary

The study evaluated cardiorespiratory deconditioning induced by head-down bed rest in late middle-aged adults. Twenty-two healthy participants, aged 55 to 65 years, were included in the study. Participants wore the Hexoskin smart t-shirt to record a one-lead electrocardiogram during incremental cycle ergometer tests. The analysis included 68 of 88 planned test records. Baseline correlation was r = 0.90 (P < 0.001) for VO2 at HRV0.75 vs VT1 and r = 0.93 (P < 0.001) for HRV0.5 vs VT2.

HexoskinConference PaperNeuroscience, Sleep & CognitionValidation & MethodologyData Science & Artificial Intelligence

HexoskinConference AbstractOccupational Health & Human FactorsClinical Medicine & Surgery

Key FindingIn a neonatal airway simulation using Hexoskin, 70% of post-scenario responses indicated stress, but residents felt stress had affected their performance only a quarter of the time.

Summary

Emergency medicine residents wore Hexoskin vests during three neonatal intubation simulations designed to become progressively more stressful. The researchers compared continuous heart and breathing measurements with participants’ ratings, observer reports and procedural performance. The poster described heart-rate spikes in all 14 participants, although their own accounts of stress did not always match observers’ assessments.

HexoskinArticleOccupational Health & Human FactorsClinical Medicine & Surgery

Key FindingThe median LF/HF ratio derived from Hexoskin cardiac recordings increased from 2.29 in the first scenario to 4.70 and 4.63 in the second and third scenarios (P=.04 for each comparison). Physiological changes occurred even when participants did not report corresponding stress, and procedural proficiency declined in the more stressful scenarios. The 12-person simulation study does not establish effects during live neonatal care.

Summary

This single-center simulation study examined subjective and physiological stress across three increasingly stressful neonatal-resuscitation scenarios. Twelve emergency-medicine residents wore Hexoskin shirts continuously from the initial survey through the final debriefing. Heart-rate variability served as an objective stress marker, while surveys assessed perceived stress and affect and intubation success represented procedural performance. Each participant served as their own baseline control.

HexoskinArticleNeuroscience, Sleep & CognitionClinical Medicine & Surgery

Key FindingIn a balance study using Hexoskin cardiac monitoring, mean Timed Up and Go time increased after the stressor from 7.00 ± 1.16 s to 7.37 ± 1.40 s (p < 0.001); without a control group, the study cannot establish that stress caused the change.

Summary

This cross-sectional study tested whether an acute emotional stressor changed fast-paced Timed Up and Go performance in older adults. The study included 31 participants. The analysis assessed TUG time before and after stressful IAPS image exposure. The Hexoskin Pro vest was worn as a monitoring tool to record the physiological data used for HRV calculation before and during viewing of stressful images, to confirm stress induction. The study reported HF (ln) 3.54 ± 0.5 pre-stress vs 3.19 ± 0.61 stress, t(30) = 3.64, p = 0.001, d = 0.653.

HexoskinConference PaperOccupational Health & Human Factors

HexoskinConference PaperOccupational Health & Human FactorsData Science & Artificial Intelligence

HexoskinAcademic ThesisData Science & Artificial IntelligenceRespiratory & Infectious DiseaseNeuroscience, Sleep & Cognition

Biomedical Signal Processing and Inference in Wearable Sensing Applications

Key FindingThe Hexoskin component demonstrated model comparison on a single healthy participant during an awake-to-sleep transition. It did not validate biometric recognition or sleep-apnea detection in a clinical population.

Summary

This thesis developed methods for ECG-based biometric recognition, cardiorespiratory modeling, and sleep-apnea inference in wearable sensing applications. Hexoskin recorded one healthy 25-year-old man during an awake-to-sleep transition to compare real physiology with cardiorespiratory-model simulations. Separate biometric analyses used the MIT-BIH Normal Sinus Rhythm database rather than Hexoskin recordings. The garment therefore served as a limited physiological data source within a broader methodological dissertation.

HexoskinArticleExercise Physiology & Sports ScienceRespiratory & Infectious Disease

Key FindingResting respiratory rate measured with Hexoskin increased from 5.1 ± 0.4 breaths/min before altitude exposure to 9.1 ± 1.6 in week 1 and 9.3 ± 2.1 in week 2; this was a single-athlete case study.

Summary

This single-athlete case study followed hematological and ventilatory responses during a five-week stay at 3,900 m. Hexoskin measured resting respiratory rate during five minutes of supine recording after awakening. Respiratory rate increased from 5.1 ± 0.4 breaths/min before altitude exposure to 9.1 ± 1.6 in week 1 and 9.3 ± 2.1 in week 2. Results from one elite wheelchair marathoner do not establish a general altitude response.

HexoskinArticleNeuroscience, Sleep & CognitionData Science & Artificial IntelligenceValidation & Methodology

Key FindingIn controlled speech tasks involving 56 adults, a model using only the thoracic respiratory signal from Hexoskin reached a test-set area under the ROC curve of 0.991, compared with 0.985 using both respiratory bands. A separate sternal accelerometer reached 0.996; performance in everyday conversation remains untested.

Summary

This comparative laboratory study trained speech-detection models using wearable data from 56 adults. Hexoskin Pro shirts supplied thoracic and abdominal respiratory inductance plethysmography signals for the dual-band and thorax-only models. Test-set AUC was 98.5% for dual-band RIP, 99.1% for thorax-only RIP, 97.8% for impedance, and 99.6% for sternal accelerometry. The controlled speech tasks do not establish performance in daily life.

AstroskinArticleExercise Physiology & Sports Science

Key FindingThe review identified 22 smart-clothing types across 24 included studies, with heart rate and respiratory rate among the commonly monitored signals. Astroskin was discussed as a commercial system rather than tested in a new validation study. Study-quality scores in the review ranged from two to five Newcastle-Ottawa stars.

Summary

This systematic review examined smart clothing used in sports and its possible application to extreme sports. The authors screened 837 records after duplicate removal and included 24 studies covering 18 systems and 22 garment types. Astroskin was summarized as one of the available products and was described as measuring blood pressure, pulse oximetry, three-lead ECG, respiration, skin temperature and activity. The review did not collect new participant data with Astroskin.

HexoskinArticleNeuroscience, Sleep & CognitionCardiology & Arrhythmia

Key FindingIn a study that used Hexoskin to monitor cardiac responses during stress, group-by-time interactions were reported for SDRR, pNN50 and LF/HF, while heart rate and cortisol did not show significant interactions and RMSSD showed a trend. The laboratory study tested short-term physiological and task responses in healthy volunteers and does not establish clinical stress-management benefit.

Summary

This randomized study compared transcranial stimulation, cardiac biofeedback and their combination during an experimental stress protocol. Eighty healthy adults completed the analysis after five of 85 enrolled participants were excluded. Hexoskin continuously tracked cardiac activity during the second session, which included baseline, anticipatory stress, a 15-minute intervention, the Trier Social Stress Test and cognitive tasks. Five-minute segments were used for heart-rate-variability analysis.

HexoskinArticleOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

Key FindingIn a simulation study that recorded cardiac activity with Hexoskin, higher neuroticism was associated with greater psychological and cardiac anticipatory stress, while extraversion was associated with lower values on both measures. Openness interacted with relaxation breathing and biofeedback in the cardiac-coherence models. These associations come from a simulation setting and do not establish that personality traits cause clinical performance differences.

Summary

This simulation study examined personality, anticipatory stress and coping responses in anesthesiology and critical-care residents. Hexoskin continuously monitored cardiac activity during anticipation, a coping intervention, a critical-care scenario, debriefing and final rest. The study enrolled 147 participants; 120 were included in anticipatory-stress models and 118 in cardiac-coherence models. Regression models tested Big Five traits and interactions with relaxation breathing, biofeedback or control conditions.

HexoskinArticleNeuroscience, Sleep & CognitionValidation & Methodology

Key FindingECG-derived respiratory rate differed from the Hexoskin respiratory-inductance reference by 17.22% on average (3.54 breaths/min); only four volunteers were studied.

Summary

This feasibility study adapted cardiopulmonary-coupling analysis for a smartphone using overnight Hexoskin recordings. Four volunteers contributed 12 whole-night recordings; the shirt streamed ECG to the app and its respiratory-inductance signal served as the reference for ECG-derived respiration. ECG-derived respiratory rate differed from the shirt reference by 17.22% on average, or 3.54 breaths/min. The four-person feasibility study does not establish clinical sleep assessment.

HexoskinConference PaperCardiology & ArrhythmiaValidation & Methodology

HexoskinAcademic ThesisExercise Physiology & Sports ScienceCardiology & Arrhythmia

Heart Rate Variability Compared in a Trained and Untrained Population under Normoxic and Hypoxic Conditions

Key FindingAmong trained participants, exercise SDNN was lower in hypoxia than normoxia only at the study's exploratory 0.10 significance level, while untrained participants showed a significant heart-rate increase in hypoxia. The sample of 13 men was too small to establish general training-group effects.

Summary

This thesis compared heart-rate variability in seven aerobically trained and six untrained men under normoxic and hypoxic conditions. Participants completed two randomized crossover visits with one hour of rest, ten minutes of submaximal exercise, and a five-minute cool-down. Hexoskin recorded ECG for SDNN calculation as well as heart rate during the protocol. The small study also compared blood pressure and oxygen saturation responses between training groups and conditions.

HexoskinConference PaperOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

HexoskinConference PaperOccupational Health & Human Factors

HexoskinConference PaperOccupational Health & Human FactorsData Science & Artificial Intelligence

HexoskinAcademic ThesisValidation & MethodologyRespiratory & Infectious Disease

A Method for Measuring Puffing and Respiratory Parameters of Tobacco Users within their Natural Usage Environment

Key FindingMean inhalation volume differed by about 27 mL between Hexoskin and the spirometer, approximately 4.5% of a roughly 600 mL mean inhalation volume. The small validation samples, long-term calibration variability, and mixed natural-use dataset limit broader accuracy claims.

Summary

This thesis developed a method for measuring tobacco-puffing and respiratory patterns in natural-use settings. Hexoskin respiratory signals were calibrated to volume, compared with a spirometer, and synchronized with a wearable puffing topography monitor. The natural-environment deployment enrolled ten tobacco users and analyzed eight datasets from seven participants across cigarette and hookah use. The work evaluated measurement and synchronization methods rather than a clinical outcome.

HexoskinArticleOccupational Health & Human FactorsExercise Physiology & Sports Science

A Mixed Method Approach to Assessing the Feasibility of Implementing the Ottawa Paramedic Physical Abilities Test (TM) in the Australian Context

Key FindingIn an Australian feasibility study using Hexoskin during a paramedic fitness assessment, six of 12 recruits completed all physical tests and two met the pass or satisfactory standard on every test. Two recruits did not start because their resting vital signs exceeded the safety thresholds.

Summary

This mixed-method feasibility study assessed whether the Ottawa Paramedic Physical Abilities Test could be implemented in Australia. Twelve paramedicine students were recruited; two did not begin testing because resting vital signs exceeded safety thresholds, six completed all physical tests, and two met every pass or satisfactory standard. The analysis used feasibility outcomes in three groups: participant characteristics and test performance, feasibility of the tests for the Australian paramedicine context, and feasibility of Hexoskin monitoring. The Hexoskin Smart Garment with its datalogger was used as a monitoring tool to record heart rate, accelerometry and ventilation during the fitness tests. The study reported 6 of 12 participants completed all physical tests and 2 completed all to a pass or satisfactory standard.

HexoskinArticleOccupational Health & Human FactorsClinical Medicine & Surgery

Key FindingFor qualified surgeons wearing Hexoskin, normalized LF/HF was 0.48 with robotic versus 1.76 with laparoscopic surgery during right gastric artery dissection (P = 0.030).

Summary

This single-center prospective cohort study examined how surgical approach and experience affected stress during minimally invasive distal gastrectomy. Four surgeons performed robot-assisted or laparoscopic procedures in 40 patients, while Hexoskin continuously recorded intraoperative heart rate variability. Stress was assessed with the LF/HF ratio across 10 standardized surgical steps, and postoperative workload was measured with NASA-TLX. Among qualified surgeons, robotic surgery was associated with lower stress than laparoscopy, especially during suprapancreatic lymph-node dissection.

HexoskinArticleValidation & MethodologyData Science & Artificial Intelligence

Key FindingThe radar method achieved 98%–100% heart-rate and respiratory-rate accuracy within 2 beats per minute up to 2 m against the Hexoskin reference; only four participants were tested.

Summary

This controlled study tested a radar method for estimating heart rate and respiratory rate in four participants. Hexoskin supplied the reference heart-rate and respiratory-rate measurements. The radar method used resonance sparse spectrum decomposition and harmonic processing to reduce clutter, motion, and respiratory interference. Accuracy was reported as 98%–100% within 2 beats per minute at distances up to 2 m, but the four-person laboratory study limits generalizability.

HexoskinConference AbstractOccupational Health & Human Factors

HexoskinArticleOccupational Health & Human Factors

Key FindingMean heart rate measured with Hexoskin increased from 81 beats/min at baseline to 99 during intubation and 108 during central-line placement, while SD1/SD2 decreased from 0.244 to 0.180 and 0.167. The five-person pilot suggests measurable physiological stress during procedures but does not establish general performance or patient-outcome effects.

Summary

This prospective pilot study monitored five pediatric critical-care physicians during live patient care. Hexoskin recorded heart rate, respiratory rate and heart-rate variability during baseline rest, patient rounds, tracheal intubation and central-line placement. Each participant was compared with their own baseline. The shirt was not compared with a reference device.

HexoskinConference AbstractExercise Physiology & Sports Science

A Comparison of Pickleball and Walking: A Pilot Study

HexoskinConference PaperValidation & MethodologyCardiology & Arrhythmia

HexoskinArticleValidation & MethodologyExercise Physiology & Sports ScienceOccupational Health & Human Factors

Key FindingDiscrepancies between Hexoskin and reference heart-rate and respiratory-rate measurements were below 10% across rest, submaximal and maximal stages, while tidal-volume and ventilation discrepancies varied by intensity. A sex- and body-weight-adjusted ventilation model reached R²=0.89 with RMSE 9.8 L/min. The healthy laboratory sample and activity-specific adjustment models do not establish accuracy in patients or unrestricted daily activity.

Summary

This laboratory study validated Hexoskin respiratory and cardiac measures during seated rest and incremental stationary cycling. Nineteen healthy bicycle commuters were tested; respiratory analyses included 17 participants and heart-rate analysis included 18. Respiratory rate, tidal volume and minute ventilation were compared with spirometry, and heart rate was compared with 12-lead ECG. Adjustment models for field use were also developed.

HexoskinArticleOccupational Health & Human FactorsCardiology & Arrhythmia

Key FindingA one-standard-deviation decrease in outdoor temperature was associated with a 0.241-log-unit increase in pNN50, or approximately 27%. RMSSD declined approximately 22% over the 49-day period. The reported temperature associations were pooled across all ECG devices and were not Hexoskin-specific.

Summary

This study investigated cardiac autonomic function during an Antarctic summer expedition at James Ross Island. The study included twelve multinational expedition members, consisting of eight males and four females with a mean age of 36.2 ± 8.7 years. The Hexoskin Pro Smart Shirt served as one of three wearable ECG devices used to record single-lead ECG over 12-hour daytime windows. Data were collected during 12-hour daytime recordings every other day over a 49-day expedition, with 45 measurement days available. Results were pooled across all ECG devices, showing that a one SD decrease in outdoor temperature was associated with a +0.241 log-units pNN50.

HexoskinArticleNeuroscience, Sleep & CognitionExercise Physiology & Sports Science

Key FindingThe rise from resting to peak heart rate measured with Hexoskin was greater during first-person-shooter play than multiplayer-online-battle-arena play (p = 0.043); the 17-person observational comparison does not establish causality.

Summary

This observational study measured physiological and cognitive changes during one competitive esports session in 17 participants. Nine played a first-person shooter and eight played a multiplayer online battle arena game, while Hexoskin monitored heart rate throughout approximately 2.5 hours of play. The rise from resting to peak heart rate was greater in the first-person-shooter group (p = 0.043). The small nonrandomized groups do not establish that game type caused the difference.

HexoskinConference AbstractExercise Physiology & Sports Science

HexoskinPreprintData Science & Artificial IntelligenceCardiology & Arrhythmia

HexoskinArticleNeuroscience, Sleep & CognitionData Science & Artificial IntelligenceClinical Medicine & Surgery

Key FindingAt a firing power threshold of 0.65, a seizure-detection system incorporating Hexoskin recordings achieved a mean sensitivity of 66% and a false alarm rate of 30.6 per 24 hours. The results include a subset of 13 out of 27 patients who demonstrated 100% event sensitivity.

Summary

The study evaluated a shirt-based detection approach for seizures annotated on video-EEG blinded to shirt data. The participants included adult patients aged 19 to 61 years who experienced focal impaired awareness seizures. The Hexoskin biometric shirt was worn continuously by epilepsy monitoring unit patients to record respiration, ECG, and accelerometry. Features from respiration, ECG HRV, and acceleration were used with an XGBoost classifier and nested leave-one-subject-out cross-validation. At a firing power threshold of 0.65, the system detected 71 seizures with a mean sensitivity of 66%.

HexoskinArticleNeuroscience, Sleep & CognitionClinical Medicine & Surgery

Key FindingParticipants completed 1,269 pre-post survey pairs, and 947 post-surveys had valid Hexoskin data; wearable data were missing for 22% of completed post-surveys on average. RMSSD, sample entropy and DFA α1 were moderately to strongly correlated, while the study's regulation findings remain observational and specific to the monitored cohort.

Summary

This ambulatory study examined autonomic complexity and affect regulation in 37 young adults with remitted major depressive disorder and 28 healthy comparison participants. Hexoskin shirts recorded single-lead ECG and acceleration during waking hours for seven days, paired with six daily ecological momentary assessments. Heart-rate-variability metrics were calculated from five-minute segments within the 30 minutes before each post-survey. The analysis tested within-person changes in complexity during affect-regulation episodes and differences between groups.

AstroskinPoster / PresentationValidation & MethodologyNeuroscience, Sleep & CognitionRespiratory & Infectious Disease

Key FindingSkin temperature correlated with the thermometer (r = 0.997), and oxygen saturation correlated with pulse oximetry (r = 0.943). The poster did not establish how accurately the system identifies individual breathing events.

Summary

This poster evaluated Astroskin signals for a study of sleep-disordered breathing after spinal cord injury. Researchers compared skin temperature and blood oxygen saturation with a digital thermometer and pulse oximeter, respectively. The poster distinguishes these sensor comparisons from its overnight recordings.

AstroskinConference AbstractNeuroscience, Sleep & CognitionValidation & MethodologyRespiratory & Infectious Disease

Key FindingIn healthy controls, Astroskin oxygen saturation had a reported bias of −2.95 percentage points relative to the reference, and skin temperature had a bias of −0.3 °C. Dynamic blood-pressure changes were captured poorly. These comparisons do not measure the ability to diagnose sleep-disordered breathing.

Summary

Researchers reported home sleep recordings from four people with spinal cord injury and four healthy controls. The abstract presents the sensor comparisons in healthy controls separately from the overnight case series.

HexoskinAcademic ThesisValidation & MethodologyClinical Medicine & Surgery

Key FindingThe proof-of-concept video signal matched the Hexoskin ECG frequency in one brief, motionless recording. That result supports signal extraction under the tested conditions but does not validate continuous pulse monitoring across patients, motion, or sleep.

Summary

This thesis developed video-based methods for non-contact monitoring of movement, respiration, and pulse in bed and clinical settings. Several experiments used healthy adults or older hospital patients, while one proof-of-concept recording used Hexoskin ECG as the reference for pulse signals extracted from visible and thermal video. The Hexoskin experiment involved one healthy volunteer who remained still for about 90 seconds, with ten seconds used in the comparison. Both the processed video region and Hexoskin ECG showed a dominant frequency of 0.9 Hz, or about 54 beats per minute.

HexoskinConference PaperData Science & Artificial IntelligenceNeuroscience, Sleep & Cognition

HexoskinArticleValidation & Methodology

Key FindingThe wireless DigiAID power-control algorithm reduced node energy use by 3%–90%; the Hexoskin comparison concerned the separate wired network and used activity-specific normal ranges rather than a clinical reference.

Summary

This study compared an experimental wired DigiAID body-sensor network with Hexoskin and then evaluated a wireless DigiAID design. Ten young adults wore DigiAID and Hexoskin simultaneously during sitting, walking, and sleeping, with Hexoskin serving as the commercial comparator for heart and respiratory rates. The dynamic power-control algorithm reduced node energy use by 3%–90% while maintaining packet reception. The paper does not state whether the separate wired and wireless participant groups were the same people, and its range-based comparisons do not establish clinical accuracy.

HexoskinAcademic ThesisOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

Police Decision-Making: The Impact of Choice on Use-of-Force Decisions

Key FindingThe thesis reported no consistent pattern of physiological arousal across the choice conditions. Because the study used a training simulation and heart rate as an arousal proxy, it does not establish how choice structure affects real-world use-of-force outcomes.

Summary

This thesis investigated whether the number of available response choices influenced use-of-force decisions by Royal Canadian Mounted Police cadets. A total of 112 cadets were analyzed after three of 115 collected records were excluded for technical issues. Hexoskin measured heart rate as an index of physiological arousal during a high-threat judgment simulation, alongside performance, response time, and confidence measures. The design compared two limited-choice groups with one extensive-choice group.

HexoskinAcademic ThesisExercise Physiology & Sports ScienceCardiology & Arrhythmia

Influence of Gestational Exercise on One-Month Infant Resting Autonomic and Cardiac Control

Key FindingExercise-group intensity predicted infant heart rate, and gestational age predicted RMSSD, with both associations reported at p = .03. Nineteen recordings were excluded because the Hexoskin system could not interpret the data, and the thesis does not establish a clinical benefit for infants.

Summary

This thesis examined resting autonomic and cardiac measures in one-month-old infants whose mothers had been assigned to different exercise programs during pregnancy. Hexoskin recorded infant heart rate, heart-rate variability, and breathing rate for at least ten minutes at the one-month visit. Seventy-four mother-infant pairs were analyzed from 95 who returned for the visit. Regression analyses related infant outcomes to the four gestational exercise groups and participant characteristics.

HexoskinArticleRespiratory & Infectious DiseaseClinical Medicine & SurgeryValidation & Methodology

Key FindingContinuous Hexoskin monitoring was feasible and generally well tolerated in five participants, but the study did not calibrate the shirt against spirometry or establish measurement accuracy.

Summary

This exploratory feasibility study assessed Hexoskin monitoring during a standardized walk with stair climbing in adolescents with idiopathic scoliosis. Five participants completed 31–35 minutes of walking while the shirt recorded tidal volume, breathing rate, minute ventilation, and heart rate. Comfort scores ranged from 6 to 9 out of 10. The shirt was not calibrated against spirometry, and the five-person descriptive study does not establish measurement accuracy.

HexoskinArticleCardiology & ArrhythmiaOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

Key FindingThe standard deviation of the LF:HF ratio derived from Hexoskin ECG during training was lower in GAD-7 screen-positive cadets than in screen-negative cadets, 1.28 versus 2.58 (P=.035), while the average LF:HF ratio did not differ. More than half of the recordings were excluded for signal artifact, and the article does not explicitly identify the 107 cadets in the comparison tables as the full analyzed cohort.

Summary

This longitudinal study assessed cardiac-autonomic variability during the 26-week Royal Canadian Mounted Police Cadet Training Program. Hexoskin garments recorded daily three-lead ECG, and 31 heart-rate-variability parameters were calculated from resting recordings in Kubios. A total of 157 cadets participated, while the anxiety-group tables reported 95 screen-negative and 12 screen-positive cadets. Of 6,699 collected recordings, 2,871 met the analysis criteria.

HexoskinConference AbstractValidation & MethodologyExercise Physiology & Sports Science

Eine explorative Studie zur Validität der Herzfrequenzmessung mittels Smart Shirts bei Laufsportlern

HexoskinReportRespiratory & Infectious DiseaseOccupational Health & Human FactorsData Science & Artificial Intelligence

Performance monitoring by simultaneous sensing of internal gas exchange and respiration patterns - Pythia phase II

HexoskinAcademic ThesisExercise Physiology & Sports ScienceData Science & Artificial IntelligenceValidation & Methodology

Prediction of Oxygen Uptake and its Dynamics by Wearable Sensors During Activities of Daily Living

Key FindingOne daily-activity analysis reported r = 0.87 between predicted and measured oxygen uptake, while another reported r = 0.69. The studies used small, mostly young and male cohorts, and three additional participants were excluded because of respiratory-band or ECG-electrode problems.

Summary

This thesis investigated wearable-sensor prediction of oxygen uptake and oxygen-uptake dynamics during walking and activities of daily living across several separate studies. In the Hexoskin studies, acceleration, ECG, and respiratory signals supplied heart rate, ventilation, breathing frequency, and movement features for machine-learning models, while gas analysis provided the oxygen-uptake reference. One unsupervised study analyzed 13 healthy men over up to four days of daily activity, with sensors worn 6.3 ± 1.4 hours per day. Predicted and measured oxygen uptake were significantly correlated in the reported analyses.

HexoskinArticleCardiology & ArrhythmiaData Science & Artificial IntelligenceValidation & Methodology

Key FindingOn the Hexoskin dataset, MS-QI correlated with actimetry at Pearson ρ=-0.94 and Spearman ρ=-0.92, outperforming the reported Pearson correlations for the comparator indices. The Hexoskin participant count and protocol were limited, and actimetry was used as the quality reference rather than expert clinical annotation.

Summary

This methods study proposed a modulation-spectrum ECG quality index for telehealth recordings. Hexoskin supplied single-channel textile ECG and actimetry from three users during sitting, walking and running, alongside synthetic and public ECG datasets. The proposed MS-QI was compared with kurtosis and an in-band-to-out-of-band power ratio. Hexoskin was a data source for algorithm evaluation rather than the device under evaluation.

AstroskinConference PaperOccupational Health & Human FactorsValidation & Methodology

Wearable Biosensor Monitor to Support Autonomous Crew Health and Readiness to Perform

Key FindingExercise signals were usable in most cases. Storing data on the Astroskin module supported 24-hour monitoring; real-time Bluetooth streaming shortened the recording time.

Summary

Researchers tested an Astroskin prototype during 30-day simulated missions in NASA’s Human Exploration Research Analog. They assessed whether crew members could record usable physiological signals during exercise and whether monitoring could continue for 24 hours.

HexoskinConference PaperExercise Physiology & Sports ScienceValidation & Methodology

HexoskinConference AbstractRespiratory & Infectious DiseaseValidation & Methodology

HexoskinConference PaperExercise Physiology & Sports ScienceOccupational Health & Human Factors

HexoskinArticleExercise Physiology & Sports ScienceOccupational Health & Human Factors

Key FindingWith respiration tracked by Hexoskin, mean absolute percentage error for breathing-rate stabilization was 1.16 ± 1.05% in study 1 and 1.9 ± 0.11% in study 2. Participants stayed attached to the instructed breathing rate for 86.81 ± 9.71% and 86.18 ± 11.96% of the time. All sounds scored below 5/10 on intention to use.

Summary

Two exploratory within-participant studies evaluated breathing-instruction sounds during treadmill running. Study 1 and study 2 each included 11 novice female runners. Hexoskin recorded respiration, heart rate, and acceleration to assess adherence to the breathing sounds. In study 1, each of five sound conditions included 5 minutes of running after a 10-minute warm-up. The small samples of female runners tested on a laboratory treadmill limit generalizability.

HexoskinConference AbstractNeuroscience, Sleep & CognitionClinical Medicine & Surgery

HexoskinArticleNeuroscience, Sleep & CognitionClinical Medicine & Surgery

Key FindingThe publication reports tracer-distribution and urinary-excretion results but no Hexoskin physiological outcomes. The methods describe subgroup counts totaling 13 rather than the stated 15 volunteers, and the article does not reconcile the remaining two participants.

Summary

This imaging study followed intrathecal tracer movement in 15 healthy volunteers using serial planar imaging and SPECT/CT. Hexoskin monitored heart rate, ventilation and physical activity, and its accelerometer was used after the study to check compliance with the physical-activity limit. The smart shirt was a monitoring aid rather than the intervention or primary measurement system. Hexoskin wear duration was not reported.

AstroskinConference PaperValidation & MethodologyExercise Physiology & Sports ScienceOccupational Health & Human Factors

Key FindingHeart-rate readings were generally highly correlated. During one participant’s treadmill test, Astroskin tracked the DADiSP result but diverged from Polar, so the overall correlation did not hold for every comparison.

Summary

Four participants exercised on a treadmill, a cycle and a rowing machine while researchers evaluated an Astroskin prototype. The team compared its heart-rate readings with a Polar monitor and with heart rate calculated in DADiSP.

HexoskinArticleValidation & MethodologyExercise Physiology & Sports Science

Key FindingHexoskin heart rate and breathing rate showed intraclass correlations of 0.96–0.99 and 0.96–1.00, respectively, with narrow reported limits of agreement. Hip-motion intensity showed intraclass correlations of 0.80–0.97, while tidal volume and minute ventilation tracked reference changes in arbitrary units. Results from healthy young volunteers and short laboratory activities do not establish clinical or free-living performance.

Summary

This two-day laboratory study validated Hexoskin during lying, sitting, standing and walking. Twenty healthy adults enrolled, although some analyses included 16, 18 or 19 participants after exclusions. Heart rate, breathing rate, tidal volume, minute ventilation and hip-motion intensity were compared with laboratory references. The final minute of each steady-state activity trial was analyzed.

HexoskinAcademic ThesisData Science & Artificial IntelligenceExercise Physiology & Sports Science

Mapper l'effort physique grâce au Hexoskin

Key FindingThe project demonstrated a software workflow for combining mapped activity with Hexoskin pulse data. Without a reported validation cohort or performance metrics, it should not be presented as evidence of physiological accuracy or training benefit.

Summary

This thesis developed an Android and web application for mapping physical effort during sports sessions. Smartphone GPS supplied route and movement information, while the web application retrieved Hexoskin data through the REST API and linked average pulse and pulse heat maps to the route. The interface also displayed duration, distance, speed, calories, and session comparisons. The thesis describes a software implementation and does not report a participant cohort or physiological validation study.

HexoskinConference PaperValidation & MethodologyData Science & Artificial IntelligenceCardiology & Arrhythmia

HexoskinPreprintData Science & Artificial IntelligenceClinical Medicine & Surgery

Towards Identifying the Optimal Timing for Near Real-Time Smoking Interventions Using Commercial Wearable Devices

HexoskinArticleData Science & Artificial IntelligenceNeuroscience, Sleep & Cognition

Key FindingIn a cognitive-state classification study using Hexoskin signals, only three of four analyzed participants showed reliably better-than-chance classification; mean accuracy was 61.3% for random forest and 57% for k-nearest neighbors.

Summary

This preliminary study tested personalized prediction of self-labelled cognitive states from Hexoskin physiological and movement data. Five people were recorded during daily work, four were included in classification, and only three showed reliably better-than-chance performance. Random-forest accuracy averaged 61.3% and k-nearest-neighbor accuracy 57% with 10-minute look-back windows. The proof-of-concept sample and self-labelled outcomes do not establish general or prospective performance.

HexoskinArticleOccupational Health & Human FactorsData Science & Artificial Intelligence

Key FindingThe lifting treatments produced mean heart rates from 134.55 to 159.12 beats/min and mean breathing rates from 18.88 to 26.20 breaths/min. The physiological values were descriptive inputs to the manufacturing-analysis framework, so the study does not establish diagnostic, fatigue-prediction or device-validation performance.

Summary

This manufacturing-methods study combined optical motion capture with wearable physiological monitoring to analyze repetitive work. A worker wore a Hexoskin bioshirt during demonstration tasks, while a separate lifting experiment involved 11 male participants. The shirt recorded heart rate, ventilation and breathing rate each second alongside motion data. The work focused on motion segmentation, dynamic-time-warping patterns and statistical process control rather than validating the garment.

HexoskinStudy ProtocolOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

HexoskinArticleExercise Physiology & Sports ScienceOccupational Health & Human Factors

Key FindingPilots monitored with Hexoskin had a higher mean heart rate at extreme than at moderate altitude (112 vs 98 beats/min, p = 0.048). Oxygen consumption during takeoff was measured separately by a metabolic system.

Summary

This field study described physiological responses during paragliding at moderate and extreme altitudes and during aerial manoeuvres. Hexoskin recorded ECG, respiratory movement and acceleration in four moderate-altitude pilots, two extreme-altitude pilots and two manoeuvre pilots. Flight phases were analyzed in five-minute sections, with additional flight data from 17 Flymaster users. All monitored pilots were male, and group sizes were small.

HexoskinConference AbstractExercise Physiology & Sports Science

HexoskinArticleOccupational Health & Human FactorsClinical Medicine & Surgery

Key FindingIn simulated encounters that used Hexoskin for physiological monitoring, self-reported state anxiety increased by 8 points for observers and 12 points for hot-seat participants (p = 0.021); the result comes from 12 participants in simulated encounters.

Summary

This simulation study examined objective and self-reported stress during critical physician communication encounters. Twelve participants wore Hexoskin shirts throughout a one-day course, including a 30-minute baseline and six simulated encounters. State anxiety increased by 8 points for observers and 12 points for participants in the hot-seat role, while several heart-rate-variability measures differed by role. The small simulation study did not validate Hexoskin against a reference or establish effects in clinical encounters.

HexoskinArticleOccupational Health & Human FactorsNeuroscience, Sleep & Cognition

Key FindingDuring a feasibility study that also used Hexoskin for physiological monitoring, questionnaire-based STAI scores decreased from 44.6 to 27.2 (p < 0.001). The study assessed the feasibility of measurement rather than the accuracy of the garment against a reference.

Summary

The study assessed the feasibility of using the Hexoskin smart shirt as an objective stress measure. Participants included 13 clinicians, including physicians, a nurse practitioner, and female pediatric nurses. The Hexoskin smart shirt recorded HRV continuously during an 8-hour mindfulness training day. During body scan exercises, RMSSD was 35.68 ± 3.79 compared to 21.90 ± 2.56 at baseline (p = 0.026).

HexoskinArticleClinical Medicine & SurgeryOccupational Health & Human Factors

Key FindingAmong residents monitored with Hexoskin, those with end-of-life care exposure had decreases in SDNN and RMSSD (P < 0.0001, P = 0.006).

Summary

This study describes the impact of physician end-of-life communication training on simulated stress using a novel stress marker. The study included 21 participants. The Hexoskin smart shirt was used to obtain baseline and in-simulation heart rate variability as the objective stress marker. Data were analyzed in 5-min heart rate variability epochs from a 30-min baseline before a 1-day retreat and three simulated encounters. The Hexoskin smart shirt was not evaluated against a reference.

HexoskinStudy ProtocolOccupational Health & Human FactorsValidation & Methodology

HexoskinConference AbstractClinical Medicine & SurgeryRespiratory & Infectious Disease

HexoskinConference PaperExercise Physiology & Sports Science

HexoskinConference PaperExercise Physiology & Sports Science

HexoskinArticleNeuroscience, Sleep & CognitionOccupational Health & Human Factors

Key FindingRespiratory-rate variability derived from Hexoskin breathing signals increased from four minutes before Dip & Waves events, while no heart-rate-variability measure showed a significant time trend; nine healthy drivers were studied.

Summary

This study tested whether respiratory variability from a Hexoskin shirt could help detect driver drowsiness. Nine healthy participants contributed 2,359 minutes of driving, and complex demodulation was used to derive respiratory frequency and variability. Respiratory-rate variability rose progressively from four minutes before heart-rate Dip & Waves events, while no heart-rate-variability metric showed a significant time trend. The small study used Dip & Waves as a drowsiness marker and did not test real-time warning performance.

HexoskinArticleNeuroscience, Sleep & CognitionRespiratory & Infectious DiseaseValidation & Methodology

Key FindingIn an ambulatory study that paired anxiety reports with Hexoskin respiratory monitoring, higher trait anxiety was associated with a lower sigh proportion (OR 0.80, 95% CI 0.74–0.87; P<.001), while the within-person anxiety-by-group interaction was significant (OR 1.14, 95% CI 1.03–1.26; P=.01). The protocol achieved 79.5% EMA completion, 84.1% high-quality respiratory samples and 81.1% EMA-respiratory linkage. Five enrolled participants were excluded for too few valid synchronized windows, so the findings require confirmation in larger cohorts.

Summary

This intensive longitudinal study paired ecological momentary assessment with continuous Hexoskin respiratory monitoring in daily life. Thirty-eight adults enrolled, and 33 contributed enough synchronized data for analysis across four 36-hour monitoring blocks. Breath-by-breath tidal-volume data were used to identify sighs in five-minute windows before anxiety ratings. The study examined state and trait anxiety associations while also assessing wear time, signal quality and linked-data coverage.

HexoskinConference PaperRespiratory & Infectious Disease

Fulgura Frango: Breath-Extended Harpsichord

HexoskinStudy ProtocolNeuroscience, Sleep & CognitionData Science & Artificial Intelligence